inforMED
MalfunctionNHV

DSU-H

Received Apr 4, 2018 · Event occurred Jan 25, 2018

Report 3003337893-2018-00001 · MDR key 7397667

Device

Generic name

System, Water Purification, General Medical U, Product Code: Nhv

Manufacturer

Nephros Inc

Model number

70-0285

Lot number

PI17/0342

Product problems

  • Break
  • Break

Patient

Not reported

  • No Patient Involvement
  • No Patient Involvement

Narrative

Additional Manufacturer Narrative

WHEN INITIALLY EVALUATION WITHOUT THE PRODUCT RETURN, IT WAS CONCLUDED THAT THE EVENT WAS NOT MDR REPORTABLE AND WAS LIKELY THAT NEPHROS DISINFECTION AND INSTALLATION PROCEDURES WERE NOT FOLLOWED. NONETHELESS, AFTER A COMPLETE INVESTIGATION THERE WAS INDEED FIBER BREAKS IN BOTH BUNDLES OF THE FILTER WHICH COULD ALLOW BACTERIA TO PASS THROUGH. THEREFORE, FROM THE INFORMATION GATHERED ON 03/05/2018, THIS IS AN REPORTABLE EVENT.

Description of Event or Problem

ON (B)(6) 2018 IT WAS REPORTED THAT A HEALTHCARE FACILITY RECEIVED A POSITIVE CULTURE DOWNSTREAM OF A DSU-H (70-0285) IN ONE OF THEIR ICE MACHINES.THERE WAS NO REPORTED ADVERSE EVENTS OR HEALTH ISSUES CONNECTED WITH THE POSITIVE RESULT. NEPHROS DSU-H IS AN IN-LINE WATER FILTER, CLASS II MEDICAL DEVICE THAT IS PLACED BEFORE ICE MACHINES IN HEALTH CARE FACILITIES IN ORDER TO RETAIN BACTERIA, VIRUS AND ENDOTOXIN. THE USER FACILITY ROUTINELY TEST THEIR WATER AT THEIR ICE MACHINE AND FOUND THAT THERE WAS POSITIVE BACTERIA DETECTED. FROM THE INITIAL DISCUSSION, IT IS LIKELY THAT OUR DISINFECTION AND INSTALLATION PROCEDURES WERE NOT FOLLOWED.