inforMED
MalfunctionNDR

DIMENSION VISTA®

Received Mar 26, 2018 · Event occurred Oct 4, 2017

Report 2517506-2018-00195 · MDR key 7371142

Device

Generic name

Dimension Vista® Multi 1 Sample Diluent

Catalog number

KD693 SMN 10469971

Lot number

7BDA81

Product problems

  • High Test Results
  • High Test Results

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

SIEMENS HEALTHCARE DIAGNOSTICS HAS CONFIRMED THAT DIMENSION VISTA® MULTI 1 SAMPLE DILUENT (MULTI 1 SDIL) LOT 7BDA81 MAY HAVE AN INCOMPLETE SLIT ON THE SEPTUM IN THE CAP OF THE VIAL. A DILUENT CAP WITH A DEFECTIVE SEPTUM HAS THE POTENTIAL TO CAUSE ERRONEOUSLY ELEVATED RESULTS IF A SAMPLE IS AUTODILUTED ONBOARD THE DIMENSION VISTA SYSTEM, DUE TO A REDUCED VOLUME OF DILUENT BEING PIPETTED FROM THE VIAL. MULTI 1 SAMPLE DILUENT (MULTI 1 SDIL) IS USED FOR DILUTIONS OF SAMPLES EXCEEDING THE ANALYTICAL MEASUREMENT RANGE WITH THE DIMENSION VISTA® ESTRADIOL (E2) FLEX REAGENT CARTRIDGE. URGENT MEDICAL DEVICE RECALL, (B)(4) DATED (B)(6) 2018 AND URGENT FIELD SAFETY NOTICE (B)(4) DATED (B)(6) 2018 WERE ISSUED TO CUSTOMERS WHO HAD BEEN SHIPPED DIMENSION VISTA® MULTI 1 SAMPLE DILUENT (MULTI 1 SDIL) LOT 7BDA81 TO INFORM THEM OF THE ISSUE.

Description of Event or Problem

A FALSELY ELEVATED ESTRADIOL (E2) RESULT WAS OBTAINED ON A CAP SURVEY SAMPLE ON THE DIMENSION VISTA 1500 INSTRUMENT. THE RESULT WAS SCORED AS UNACCEPTABLE BY THE CAP. THE CUSTOMER STATED THAT NO PATIENT SAMPLES HAVE BEEN AFFECTED.

Remedial action

  • Recall