inforMED
OtherMUJ

ECLIPSE

Received Jul 13, 2006 · Event occurred Jun 15, 2006

Report 3003793371-2006-00006 · MDR key 735689

Device

Generic name

Treatment Planning System

Model number

*

Catalog number

48

Lot number

*

Product problems

  • Use of Device Problem

Patient

*

  • Underdose

Narrative

Description of Event or Problem

THE CUSTOMER REPORT INDICATES THAT DURING RADIATION THERAPY TREATMENT, THE PATIENT'S TREATMENT PLAN DID NOT INCLUDE THE BOLUS, RESULTING IN 100CGY UNDER DOSE. THE PHYSICIAN REVIEWED THE PLAN AND DECIDED TO ADJUST THE DOSAGE OVER THE REMAINING 12 FRACTIONS BY CREATING A NEW PLAN, LINKING THE BOLUS WITH THE NEW DOSE/FRACTION TO COMPENSATE FOR THE PREVIOUS TREATMENTS.

Additional Manufacturer Narrative

NO MALFUNCTION OF THE DEVICE OCCURRED IN THIS INCIDENT. THIS IS THE CORRECT FUNCTIONALLY OF THE SOFTWARE. USER ERROR RELATED.