API® 20E TEST STRIP

Received Mar 16, 2018

Report 3002769706-2018-00042 · MDR key 7344955

Device

Generic name

Api® 20e Test Strip

Manufacturer

Biomerieux Sa

Catalog number

20100

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings
  • Incorrect, Inadequate or Imprecise Result or Readings

Patient

Not reported

  • No Patient Involvement
  • No Patient Involvement

Narrative

Description of Event or Problem

A CUSTOMER IN (B)(6) CONTACTED BIOMÉRIEUX TO REPORT THE OCCURRENCE OF A MISIDENTIFICATION OF RAOULTELLA PLANTICOLA (EXTERNAL QC STRAIN) AS KLEBSIELLA PNEUMONIA IN ASSOCIATION WITH THE API® 20E TEST. NO ALTERNATE TEST METHOD OR CONFIRMATORY TESTING WAS PERFORMED BY THE CUSTOMER. THE EXTERNAL QC TESTING WAS PERFORMED ONE YEAR AGO; THE STRAIN IS NO LONGER AVAILABLE FOR RETEST. PRIOR TO 2001, THE NAME OF RAOULTELLA PLANTICOLA WAS KLEBSIELLA PLANTICOLA. THE BIOCHEMICAL CHARACTERISTICS BETWEEN RAOULTELLA PLANTICOLA AND KLEBSIELLA PNEUMONIA ARE SIMILAR. IT IS IMPORTANT TO NOTE THAT RAOULTELLA PLANTICOLA IS NOT IN THE API® 20E DATABASE; THEREFORE, IT IS NOT POSSIBLE TO IDENTIFY RAOULTELLA PLANTICOLA THEAPI® 20E TEST STRIP. THERE IS NO INDICATION OR REPORT FROM THE LABORATORY THAT THE DISCREPANT RESULT LED TO ANY ADVERSE EVENT RELATED TO ANY PATIENT'S STATE OF HEALTH. THERE WAS NO PATIENT DIRECTLY ASSOCIATED WITH THE QC STRAIN. BIOMÉRIEUX INVESTIGATION WILL BE INITIATED.

Additional Manufacturer Narrative

DURING AN EXTERNAL QUALITY CONTROL (QC )(RING TRIAL, RFB), THE CUSTOMER TESTED API® 20 E STRIPS AND OBTAINED A "BAD IDENTIFICATION" TO "KLEBSIELLA PNEUMONIAE" AT 97.3%. THE INTENDED IDENTIFICATION WAS "RAOULTELLA PLANTICOLA." AN INTERNAL BIOMÉRIEUX INVESTIGATION WAS INITIATED; HOWEVER, THE EXTERNAL QC TOOK PLACE ONE (1) YEAR AGO SO THE STRAIN IS NO LONGER AVAILABLE. IN ADDITION, THERE WAS NO RETURN OF THE API 20 E STRIPS AND THE LOT NUMBER OF THE BATCH WAS UNKNOWN. IN THE ABSENCE OF INFORMATION, INCLUDING CUSTOMER TROUBLESHOOTING , INVESTIGATION COULD NOT BE DONE INTERNALLY. CONSEQUENTLY, IT WAS NOT POSSIBLE TO DETERMINE THE PROBABLE CAUSE OF THE ISSUE AND TO IMPLEMENT ACTION. NOTE: PRIOR TO 2001, THE NAME OF RAOULTELLA PLANTICOLA WAS "KLEBSIELLA PLANTICOLA". THE ARTICLE, "INTERNATIONAL JOURNAL OF SYSTEMATIC AND EVOLUTIONARY MICROBIOLOGY" (2001), 51, 925-932 BY M. DRANCOURT, TITLE "PHYLOGENETIC ANALYSES OF KLEBSIELLA SPECIES DELINEATE KLEBSIELLA AND RAOULTELLA GEN. NOV, WITH DESCRIPTION OF RAOULTELLA ORNITHINOLYTICA COMB. NOV., RAOULTELLA TERRIGENA COMB. NOV. AND RAOULTELLA PLANTICOLA COMB. NOV." MENTIONS: " CLUSTER II CONTAINED KLEBSIELLA ORNITHINOLYTICA, KLEBSIELLA PLANTICOLA, KLEBSIELLA TREVISANII AND KLEBSIELLA TERRIGENA, ORGANISMS CHARACTERIZED BY GROWTH AT 10°C AND UTILIZATION OF L-SORBOSE AS CARBON SOURCE. [...] THE NAME RAOULTELLA AS A GENUS NAME FOR SPECIES OF CLUSTER II AND EMENDED DEFINITIONS OF KLEBSIELLA SPECIES ARE PROPOSED." IN ADDITION, THE "RAOULTELLA PLANTICOLA" IS NOT IN THE DATABASE OF THE API 20 E STRIPS, SO IT IS NOT IDENTIFIABLE WITH THESE STRIPS. API 20 E STRIPS GAVE IDENTIFICATION TO "KLEBSIELLA PNEUMONIAE" BECAUSE IT IS A CLOSER STRAIN TO "KLEBSIELLA PLANTICOLA" (="RAOULTELLA PLANTICOLA").