inforMED
MalfunctionEGN

TC TENOTOMY SCISSDELCVD1/SERR130MM

Received Mar 15, 2018 · Event occurred Aug 29, 2016

Report 9610612-2018-00100 · MDR key 7343488

Device

Generic name

Dental Scissors

Manufacturer

Aesculap Ag

Model number

DO250R

Catalog number

DO250R

Lot number

4507434213

Product problems

  • Break
  • Material Integrity Problem
  • Break
  • Material Integrity Problem

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

EXEMPTION NUMBER: E2014018. MANUFACTURING SITE EVALUATION: THE TIP OF THE SCISSORS ARE BROKEN. PICTORIAL DOCUMENTATION WAS CARRIED OUT WITH A DIGITAL MICROSCOPE. HARDNESS MEASUREMENT WAS CARRIED OUT WITH HARDNESS TESTER. THE DEVICE HISTORY FILES HAVE BEEN CHECKED AND WERE FOUND TO BE ACCORDING TO THE SPECIFICATIONS. BASED ON THE INFORMATION AVAILABLE AS WELL AS A RESULT OF THE INVESTIGATION THE ROOT CAUSE OF THE FAILURE IS MOST PROBABLY RELATED TO AN INSUFFICIENT USAGE. THIS KIND OF INSTRUMENT IS DESIGNATED FOR DELICATE USE ONLY. IT IS LIABLE THAT A MECHANICAL OVERLOAD SITUATION LED TO THE BREAKAGE. THE VISUAL INVESTIGATION AND THE HARDNESS TEST INDICATED THAT THE QUALITY REQUIREMENTS ARE IN THE SPECIFIED TOLERANCE RANGE. A CAPA IS NOT NECESSARY.

Description of Event or Problem

COUNTRY OF COMPLAINT: USA. CUSTOMER STATES THAT DURING INTRA-OPERATIVE SURGERY THE TIP OF SCISSORS BROKE OFF. NO SURGICAL DELAY AND NO PATIENT INJURY REPORTED.