ORS-100N WARMER DRAPE
Received Mar 13, 2018
Report 8043817-2018-00007 · MDR key 7335774
Device
Generic name
Fluid Warming And Slush Drapes
Manufacturer
Microtek Dominicana, S.a.Model number
ORS-100NLot number
D161731
Product problems
- Hole In Material
- Hole In Material
Patient
Not reported
- No Patient Involvement
- No Patient Involvement
Narrative
Additional Manufacturer Narrative
THE SAMPLE RECEIVED WAS EVALUATED. A SMALL CUT/HOLE WAS OBSERVED ABOUT 5 1/2 INCHES FROM THE CENTER OF THE DRAPE. THIS LOOKS TO HAVE BEEN CAUSED BY A SHARP OF SOME SORT. THE DHR WAS REVIEWED AND IT WAS NOTICED THAT THIS LOT HAD 95 CASES THAT WERE MANUFACTURED FROM 06/21/2016 TO 06/22/2016. NO DEFECTS WERE REPORTED DURING QUALITY INSPECTIONS. THIS IS A NON STERILE PART THAT IS SUPPLIED TO A KIT PACKER FOR FURTHER PACKAGING AND STERILIZATION. BASED ON THE DEVICE HISTORY RECORD AND SAMPLE REVIEW, THIS DOES NOT APPEAR TO BE THE RESULT OF A PERSONNEL, PROCESS OR MATERIAL ISSUE. BECAUSE THIS APPEARS TO BE RELATED TO MISUSE BY THE CUSTOMER, NO ACTIONS ARE BEING TAKEN AT THIS TIME.
Description of Event or Problem
A CUSTOMER IS REPORTING A COMPLAINT FROM A HOSPITAL STATING THAT THEY RECEIVED A FLUID WARMER DRAPE THAT CONTAINED A HOLE WITHIN THE DRAPE. THE SAMPLE WAS RECEIVED BY THE CUSTOMER AND THE REPORTED CONCERN WAS CONFIRMED. A SMALL TEAR WITHIN THE PLASTIC WAS NOTED AND UPON EXAMINING THE LOCATION OF THE DRAPE, IT WAS NOTED THAT THERE WAS NOTHING SHARP NEXT TO THE DRAPE THAT COULD HAVE CAUSED SUCH A TEAR. ALSO, THE DRAPE IS RECEIVED FOLDED AND IS PLACED WITHIN THE KIT AS SUCH.