XIA TITANIUM 4.5 VITALIUM ROD 4.5 X 600 MM

Received Mar 8, 2018 · Event occurred Feb 12, 2018

Report 0009617544-2018-00064 · MDR key 7323863

Device

Generic name

Thoracolumbosacral Pedicle Screw System.

Manufacturer

Stryker Spine-us

Model number

48132601

Catalog number

48132601

Lot number

UNKNOWN

Product problems

  • Fracture
  • Break
  • Fracture

Patient

Not reported

  • Injury
  • No Known Impact Or Consequence To Patient
  • Injury
  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE ROD BROKE.

Additional Manufacturer Narrative

THE REPORTED EVENT RESULTED IN PATIENT REVISION SURGERY. METHOD: RISK ASSESSMENT; X-RAYS. RESULTS: VISUAL, DIMENSIONAL AND FUNCTIONAL ANALYSIS COULD NOT BE PERFORMED AS THE DEVICE WAS NOT RETURNED. NO LOT # WAS PROVIDED, SO A MANUFACTURING RECORD REVIEW COULD NOT BE PERFORMED. CONCLUSION: THE RODS WERE FOUND FRACTURED ABOUT 9 MONTHS AFTER THE ORIGINAL SURGERY. FUSION DID NOT OCCUR. THE X RAY SHOWS SEVERAL LEVELS OF FIXATION AT THE TOP OF THE CONSTRUCT, AND SEVERAL LEVELS OF FIXATION AT THE BOTTOM, AND A SECTION OF NON FIXATION IN THE MIDDLE. THIS, ALONG WITH THE CURVATURE OF THE SPINE, COULD CAUSE A BENDING MOMENT. OVER TIME THIS COULD CAUSE THE RODS TO FRACTURE DUE TO THE STRESS AND STRAIN OF THE CONSTRUCT. HOWEVER SINCE THE DEVICE WAS NOT RETURNED FOR EVALUATION THE DEFINITE ROOT CAUSE CANNOT BE DETERMINED.

Description of Event or Problem

IT WAS REPORTED THAT THE ROD BROKE.