inforMED
OtherCCE

ABL800 SERIES

Received Jun 19, 2006 · Event occurred May 16, 2006

Report 3002807968-2006-00030 · MDR key 730804

Device

Generic name

*

Model number

ABL825

Catalog number

*

Lot number

*

Product problems

  • Incorrect Or Inadequate Test Results

Patient

Not reported

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

AFTER TREATMENT OF FOUR PATIENTS BASED ON MEASUREMENTS FROM THE BLOODGAS ANALYSER ABL825, IT WAS OBSERVED THAT ALL THE FOUR MEASUREMENT RESULTS USED FOR THE TREATMENT, WERE IDENTICAL WITH THE MEASUREMENT CONDUCTED JUST BEFORE THE MEASUREMENTS OF THE FOUR PATIENTS. NO HARM TO THE PATIENTS HAS BEEN REPORTED. THE PROBLEM IS CAUSED BY A SW BUG.