ELECSYS T4 ASSAY
Received Feb 27, 2018 · Event occurred Feb 8, 2018
Report 1823260-2018-00587 · MDR key 7300026
Device
Generic name
Radioimmunoassay, Total Thyroxine
Manufacturer
Roche DiagnosticsModel number
NACatalog number
ASKU
Lot number
ASKU
Product problems
- High Test Results
- High Test Results
Patient
Not reported
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
(B)(6). (B)(4).
Description of Event or Problem
THE CUSTOMER COMPLAINED OF HIGH RESULTS FOR 1 PATIENT SAMPLE TESTED FOR ELECSYS FT4 II ASSAY (FT4 II) AND ELECSYS FT3 III (FT3 III) ON A COBAS 6000 E 601 MODULE. THE INITIAL FT3 III AND FT4 II RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY WHERE THE PHYSICIAN COMPLAINED THAT THE RESULTS DID NOT CORRESPOND TO THE PATIENT'S CLINICAL PICTURE. THE SAMPLE WAS SENT FOR TESTING BY THE ABBOTT METHOD ALONG WITH REPEAT TESTING WITH AN EDTA PLASMA SAMPLE ON THE CUSTOMER'S E601 MODULE AND REPEAT TESTING ON ANOTHER HOSPITALS E601 MODULE. BASED ON THE DATA PROVIDED, DISCREPANT RESULTS WERE IDENTIFIED FOR FT4 II, FT3 III, ELECSYS T3 (T3) AND ELECSYS T4 ASSAY (T4). REFER TO MEDWATCHES WITH THE FOLLOWING PATIENT IDENTIFIERS FOR THE ASSAYS INVOLVED: (B)(6). THERE WAS NO ALLEGATION THAT AN ADVERSE EVENT OCCURRED. THE CUSTOMER'S E601 SERIAL NUMBER WAS (B)(4). THE E601 SERIAL NUMBER USED AT THE EXTERNAL LABORATORY WAS NOT PROVIDED.
Additional Manufacturer Narrative
UPON FURTHER INVESTIGATION OF THE PATIENT SAMPLE, AN INTERFERENT AGAINST A COMPONENT OF THE REAGENT WAS CONFIRMED. THIS CAUSED THE HIGH FT4 II, FT3 III, T3 AND T4 RESULTS. THIS INTERFERENCE IS COVERED IN PRODUCT LABELING. PRODUCT LABELING STATES "IN RARE CASES, INTERFERENCE DUE TO EXTREMELY HIGH TITERS OF ANTIBODIES TO ANALYTE-SPECIFIC ANTIBODIES, STREPTAVIDIN OR RUTHENIUM CAN OCCUR. THESE EFFECTS ARE MINIMIZED BY SUITABLE TEST DESIGN. FOR DIAGNOSTIC PURPOSES, THE RESULTS SHOULD ALWAYS BE ASSESSED IN CONJUNCTION WITH THE PATIENT¿S MEDICAL HISTORY, CLINICAL EXAMINATION AND OTHER FINDINGS."