ULTRACISION SURG DEV, BLADES 10MM REUSABLE, OPEN
Received Jun 19, 2006
Report 1527736-2006-02764 · MDR key 727249
Device
Generic name
Geo
Manufacturer
Ethicon Endo Surgery, Inc.(cincinnati)Model number
NACatalog number
DH010
Lot number
UNK
Product problems
- Replace
- Device Issue
Patient
UNKNOWN
- No Consequences Or Impact To Patient
Narrative
Description of Event or Problem
DURING A CHOLECYSTECTOMY PROCEDURE NO FUNCTION AFTER A FEW MINUTES. A NEW DEVICE WAS USED TO COMPLETE THE CASE WITH NO ADVERSE PT CONSEQUENCE. THERE IS NO FURTHER INFO AVAILABLE.
Additional Manufacturer Narrative
BASED ON ANALYSIS RESULTS, THIS COMPLAINT IS NOW DETERMINED TO BE A MDR MALFUNCTION. THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED WITH THE BLADE SCRATCHED AND CRACKED. THE DEVICE WAS TESTED WITH A GENERATOR AND AN ERROR CODE 5 WAS RECEIVED. THE IDENTIFIED BLADE DAMAGE MAY HAVE OCCURRED FROM EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE. IN ADDITION, MINOR BLADE DAMAGE MAY INCREASE IN SEVERITY DURING SUBSEQUENT ACTIVATIONS, AND MAY RESULT IN BLADE "LOCKOUT" LATER IN THE PROCEDURE. THEREFORE, THE INSTRUCTIONAL INSERT STATES: "SCRATCHES ON THE BLADE MAY LEAD TO PREMATURE BLADE FAILURE" AND "AVOID ACCIDENTAL CONTACT WITH OTHER INSTRUMENTS DURING USE." THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MFG PROCESS. 510(K) NUMBER IS K990362.
Remedial action
- Other