PORT INTRODUCER KIT
Received Feb 6, 2018 · Event occurred Nov 24, 2017
Report 7244264 · MDR key 7244264
Device
Generic name
Port Introducer Kit
Manufacturer
Cardinal HealthProduct problems
- Detachment Of Device Component
- Fluid/Blood Leak
- Chemical Spillage
Patient
Not reported
- Chemical Exposure
- Chemical Exposure
Narrative
Description of Event or Problem
PATIENT IN FOR CADD DISCONNECT, STATING AT 4 A.M. WHEN HE GOT UP TO USE THE BR HE FELT LIKE HIS SHIRT WAS WET AND NOTICED THAT THE TUBING FOR PORT ACCESS TO THE CHEMO BALL WAS BROKEN AND BLOOD WAS COMING OUT OF TUBING PROXIMAL TO PORT SITE. HE STATES THAT HE HAD TO CUT THE TUBING TO GET THE CLAMP TO STOP THE BLOOD FROM COMING OUT OF THE PORT. PATENT STATES THE BALL WAS FLAT AT THIS TIME AND A VERY SMALL AMOUNT OF CHEMO WAS SPILLED. PATIENT STATES THAT THEY USED THE SPILL KIT TO CLEAN UP THE SPILL. PATIENT STATES THAT HE WENT TO THE ER TO BE EVALUATED. THE ER LEFT THE PORT NEEDLE IN PLACE AND PATIENT CAME TO UNIT TO BE DE ACCESSED. PORT NEEDLE TAKEN OUT DUE TO NO PLACE FOR IT TO BE ACCESSED AND FLUSHED. PATIENT THEN RE ACCESSED AND FLUSHED PER PROTOCOL WITH NO DIFFICULTIES. PATIENT INSTRUCTED TO BAG UP HIS CLOTHING FROM THE SPILL UNTIL RN COULD FURTHER INVESTIGATE WHAT APPROPRIATE ACTION TO TAKE WITH THE CLOTHING. MD MADE AWARE, WITH NO NEW ORDERS AT THIS TIME.