ARTHROSCOPIC SHAVERS
Received Jan 15, 2018 · Event occurred Dec 29, 2017
Report 3032391-2018-00001 · MDR key 7191376
Device
Generic name
Arthrex Coolcut¿ Excalibur (blue/beige) 4.0mm
Manufacturer
Medline RenewalModel number
AR-8400EXLot number
367956
Product problems
- Device Issue
- Device Issue
Patient
16 YR
- No Consequences Or Impact To Patient
- No Consequences Or Impact To Patient
Narrative
Additional Manufacturer Narrative
THE DEVICE IN QUESTION WAS RETURNED FOR EVALUATION. IT WAS CONFIRMED THE DEVICE WAS REPROCESSED. THE LOT NUMBER ASSOCIATED WITH THE DEVICE WAS 367956. A REVIEW OF THE REPROCESSING RECORDS WAS PERFORMED. ALL PROCESSES WERE CONDUCTED AS REQUIRED PRIOR TO RELEASE. THERE WERE NO SHAVINGS ON THE RETURNED SAMPLE, FUNCTIONAL TESTS ON THE SAMPLE DETERMINED THAT THE SAMPLE FUNCTIONED ACCORDING TO PRODUCT DESIGN AND THE SAMPLE MET PRODUCT SPECIFICATIONS. WE HAVE NOT IDENTIFIED A MANUFACTURING DEFECT TO BE A CAUSE OR CONTRIBUTING FACTOR TO THE INITIAL REPORTED ISSUE. A ROOT CAUSE COULD NOT BE DETERMINED HOWEVER, DUE TO THE REPORTED INCIDENT AND THE SUBSEQUENT NEED FOR SURGICAL INTERVENTION, THE MEDWATCH IS BEING FILED.
Description of Event or Problem
WE RECEIVED A REPORT INDICATING THAT A REPROCESSED ARTHREX COOLCUT¿ EXCALIBUR SHAVER WAS BLOWING METAL CHUNKS OUT OF THE TIP DURING THE PROCEDURE. THE DEBRIS WAS REMOVED BY THE SURGEON USING IRRIGATION, ANOTHER DEVICE WAS OPENED AND THE PROCEDURE CONTINUED WITHOUT FURTHER INCIDENT.