MalfunctionKGI
DISCOVERY
Received Jan 10, 2018 · Event occurred Dec 6, 2017
Report 1220984-2018-00056 · MDR key 7180408
Device
Generic name
Densitometer, Bone
Manufacturer
Hologic, IncModel number
DISCOVERY-WCatalog number
N/A
Lot number
N/A
Product problems
- Device Displays Incorrect Message
- Device Displays Incorrect Message
Patient
Not reported
- Not Applicable
- Not Applicable
Narrative
Description of Event or Problem
IT WAS REPORTED THAT THE SYSTEM RECEIVED AN ERROR AT THE END OF SCAN WHICH PREVENTED THE SCAN FROM BEING VIEWABLE AND CAUSED THE PATIENT TO BE RE-EXPOSED. IT WAS DETERMINED THAT AN ODBC DATABASE REPAIR WAS NEEDED. ONCE THE ODBC DATABASE REPAIR WAS COMPLETED THE SYSTEM IS WORKING AS INTENDED.