COLLECT.NO.QAS KNEE IMPLANTS VEGA
Received Jan 5, 2018
Report 9610612-2018-00006 · MDR key 7169987
Device
Generic name
Knee Endoprosthetics
Manufacturer
Aesculap Implant SystemsModel number
AE-QAS-K521-56Catalog number
AE-QAS-K521-56
Product problems
- Failure To Adhere Or Bond
- Detachment Of Device Component
- Failure To Adhere Or Bond
- Detachment Of Device Component
Patient
Not reported
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
EXEMPTION NUMBER: E2014018. MANUFACTURING SITE EVALUATION: EVALUATION ON-GOING DEVICE NOT RETURNED.
Description of Event or Problem
COUNTRY OF COMPLAINT: USA. ABOUT A YEAR AGO, A SURGEON IMPLANTED A VEGA TOTAL KNEE AT (B)(6) HOSPITAL. UPON FOLLOW UP, THE VEGA TIBIAL COMPONENT HAS LOOSENED. THEY BECAME AWARE THAT THIS OCCASIONALLY HAPPENS; HOWEVER, THE SURGEON NOW HAS CONFIRMED THAT THIS IN NOT AN ISOLATED INCIDENT. HE NOW HAS TWO PATIENTS THAT HAVE CONFIRMED LOOSE TIBIA BASEPLATES (AND THEY ARE IN NEED OF REVISION). ADDITIONALLY, HE IS "KEEPING A CLOSE EYE ON" TWO MORE PATIENTS THAT HE SUSPECTS HAVE LOOSE TIBIAS. THIS IS A TOTAL OF 4 PATIENTS. ALL PATIENTS REPORTED ARE LESS THAN A YEAR POST OP. THE SURGEON SUSPECTS THAT THIS ISSUE MAY HAVE COME FROM SURGICAL TECHNIQUE. INITIALLY, HE IMPLANTED THE TIBIAL, FEMORAL, AND MENISCAL COMPONENTS INTO THE PATIENT AND LOCKED THE POLY SCREW INTO PLACE BEFORE THE CEMENT HAD A CHANCE TO HARDEN. APPARENTLY, THE SURGEON WAS INFORMED THAT THIS A IMPROPER TECHNIQUE - HE SHOULD WAIT UNTIL THE CEMENT HAS HARDENED BEFORE THE POLY SCREW IS IMPLANTED. HE HAS SINCE CHANGED HIS SURGICAL TECHNIQUE. IF THIS IS THE REASON WHY THERE ARE LOOSE TIBIA BASEPLATES IN HIS PATIENTS, THEN THE PROBLEM IS SOLVED (HOPEFULLY WE DON'T SEE ANY MORE LOOSENING). NO INJURY TO PATIENT. NO DELAY IN SURGERY. NO ADDITIONAL INTERVENTION REQUIRED. ALL MED WATCH SUBMISSIONS RELATED TO THIS REPORT ARE: 9610612-2018-00005, 9610612-2018-00006.
Additional Manufacturer Narrative
INVESTIGATION: NO PRODUCT IS AT HAND. CONCLUSION AND ROOT CAUSE: NO PRODUCT AVAILABLE AND THEREFORE AN ANALYSIS IS NOT POSSIBLE. NO CAPA IS NECESSARY.