inforMED
InjuryFSO

OPMI PENTERO 800

Received Dec 26, 2017 · Event occurred Nov 15, 2017

Report 9615010-2017-00019 · MDR key 7146459

Device

Generic name

Microscope, Surgical, General & Plastic Surgery

Model number

800

Catalog number

302582-9903-000

Lot number

NA

Product problems

  • Adverse Event Without Identified Device or Use Problem
  • Adverse Event Without Identified Device or Use Problem

Patient

Not reported

  • Burn(s)
  • Burn(s)

Narrative

Additional Manufacturer Narrative

THE DIFFERENT FACTORS THAT CAN INFLUENCE THE RISK OF BURN INJURIES ARE DESCRIBED IN DETAIL IN THE USER MANUAL (G-30-1458-EN, ISSUE 9.1). RECOMMENDATIONS FOR REDUCING THE RISK OF BURNS ARE ALSO INCLUDED.

Description of Event or Problem

THE HEALTH CARE PROFESSIONAL (HCP) REPORTED THAT A HEAD AND NECK CANCER PATIENT SUSTAINED A BURN AFTER A FLAP SURGERY IN WHICH THE OPMI PENTERO MICROSCOPE WAS USED. THE MICROSCOPE LIGHT INTENSITY SETTING WAS AT 87% AND THRESHOLD VALUE LIGHT WARNING WAS AT 81%. IN THE FIRST HALF HOUR OF THE THREE HOURS SURGERY, THE HCP RECOGNIZED THAT THE VEIN HAD DRIED ON THE PATIENT'S NECK . AN ABUNDANT IRRIGATION WAS PERFORMED WITH SERUM PHYSIOLOGY AS AN AFTER TREATMENT.