ORS-331 SLUSH WARMER DRAPE WITH PLATE

Received Dec 12, 2017

Report 8043817-2017-00033 · MDR key 7110751

Device

Generic name

Fluid Warming And Slush Drapes

Model number

ORS-331

Lot number

UNKNOWN

Product problems

  • Hole In Material
  • Hole In Material

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

A SAMPLE WAS RECEIVED FOR EVALUATION. A SMALL BURN MARK/HOLE WAS OBSERVED ABOUT 16 INCHES FROM THE SLUSH PLATE. NO LOT NUMBER WAS PROVIDED TO PERFORM A DEVICE HISTORY RECORD REVIEW. BASED ON THE SAMPLE EVALUATION THIS DOES NOT APPEAR TO BE THE RESULT OF A PERSONNEL, PROCESS, OR MATERIAL ISSUE. MELTS, THOUGH UNCOMMON, CAN OCCUR WHEN THE WARMER IS NOT DRAPED PROPERLY, NOT TURNED OFF BY USING THE POWER BUTTON, AND/OR THERE IS INSUFFICIENT FLUID IN THE WARMER BASIN, CONTRARY TO THE OPERATIONS MANUAL, PRODUCT LABELING, PRODUCT INSERT AND IN-SERVICE PRESENTATIONS. BECAUSE THIS APPEARS TO BE RELATED TO MISUSE BY THE CUSTOMER, NO ACTIONS ARE BEING TAKEN AT THIS TIME.

Description of Event or Problem

COMPANY REP REPORTED THAT (B)(6) IS STATING THAT THEY FOUND ABOUT AN INCH OF SALINE IN THE BOTTOM OF THE WARMER AFTER A SURGICAL CASE. THE HOSPITAL RECEIVED THE DRAPE THROUGH A CARDINAL KIT. NO PATIENT INJURY OR TREATMENT WAS REPORTED FOR THE INCIDENT.