EEG-1200A

Received Dec 6, 2017 · Event occurred Nov 7, 2017

Report 8030229-2017-00424 · MDR key 7087049

Device

Generic name

Neurofax Electroencephalograph

Model number

EEG-1200A

Catalog number

EEG-1200A

Product problems

  • Smoking
  • Smoking

Patient

Not reported

  • No Consequences Or Impact To Patient
  • No Known Impact Or Consequence To Patient
  • No Consequences Or Impact To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

ADDITIONAL INFORMATION: TYPE OF REPORT? IF FOLLOW-UP, WHAT TYPE? DETAILS OF COMPLAINT: CUSTOMER CALLED AND REPORTED HER EEG/POLYSMITH SYSTEM 'POPPED AND STARTED SMELLING LIKE SOMETHING WAS BURNT. UNIT WAS SENT IN FOR EVALUATION. SERVICE REQUESTED: REPAIR/LOANER. SERVICE PERFORMED: TROUBLESHOOTING. INVESTIGATION RESULT: REVIEWED THE PC WITH REPAIR CENTER AND THE COMPUTER WOULD NOT TURN ON, ONLY THE AMBER LIGHT WAS FLASHING. THERE WAS NO BURNING SMELL. REQUESTED THE REPAIR CENTER CONTINUE WITH TROUBLESHOOTING PC. QA/REPAIR CENTER HAS EVALUATED THE BURNING SMELL THAT CAME FROM POWER SUPPLY BURNING. REPLACE THE POWER SUPPLY. REAPPLY THE CPU PASTE. CHECK THE EEG TEST.

Additional Manufacturer Narrative

THE EEG TECH REPORTED THE EEG / PSG COMBO SYSTEM "POPPED AND STARTED SMELLING LIKE SOMETHING WAS BURNT." QA ALONG WITH SERVICE REVIEWED THE PC AND THE COMPUTER WOULD NOT TURN ON, ONLY THE AMBER LIGHT WAS FLASHING. THERE WAS NOT BURNING SMELL. THE UNIT WAS THEN EVALUATED BY THE SERVICE, AND THEY CONCLUDED THAT THE POWER SUPPLY WAS BAD. THE POWER SUPPLY WAS REPLACED AND THE UNIT WAS BURNED-IN FOR 72 HOURS AND EVERYTHING TESTED FINE. NIHON KOHDEN WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem

THE EEG TECH REPORTED THE EEG / PSG COMBO SYSTEM "POPPED AND STARTED SMELLING LIKE SOMETHING WAS BURNT."