MalfunctionCDX

ELECSYS T4 ASSAY

Received Nov 29, 2017 · Event occurred Mar 7, 2017

Report 1823260-2017-02772 · MDR key 7067047

Device

Generic name

Radioimmunoassay, Total Thyroxine

Manufacturer

Roche Diagnostics

Model number

NA

Catalog number

12017709122

Lot number

19428105

Product problems

  • High Test Results
  • High Test Results

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

(B)(4). THIS EVENT OCCURRED IN (B)(6).

Description of Event or Problem

THE CUSTOMER RECEIVED QUESTIONABLE RESULTS FOR MULTIPLE ASSAYS FOR ONE PATIENT FROM AN UNKNOWN ROCHE ANALYZER. THE RESULTS WERE REPORTED TO THE ENDOCRINOLOGIST. THE PATIENT WAS UNDER GROWTH HORMONE SUPPLEMENTATION THERAPY WITH SAIZEN AND THE ENDOCRINOLOGIST CLAIMED THE RESULTS DID NOT FIT THE CLINICAL PICTURE. INTERFERENCE WAS SUSPECTED. OF THE DATA PROVIDED ONLY THE RESULTS FOR ELECSYS TSH ASSAY, ELECSYS T4 ASSAY, AND ELECSYS ESTRADIOL III ASSAY WERE A REPORTABLE MALFUNCTION. REFER TO THE ATTACHMENT TO THE MEDWATCH FOR ALL PATIENT DATA. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. (B)(4).

Additional Manufacturer Narrative

SAMPLE FROM THE PATIENT WAS SUBMITTED FOR INVESTIGATION AND AN INTERFERING FACTOR TO A COMPONENT OF THE ASSAY WAS FOUND IN THE SAMPLE. PRODUCT LABELING FOR THE ASSAY DOCUMENTS THIS INTERFERENCE. THE INCIDENCE RATE OF THE IDENTIFIED INTERFERING FACTOR IS MONITORED ON A QUARTERLY BASIS.