IMMULITE/IMMULITE 1000 THYROGLOBULIN
Received Nov 17, 2017 · Event occurred Oct 26, 2017
Report 2432235-2017-00618 · MDR key 7041483
Device
Generic name
Immulite/immulite 1000 Thyroglobulin
Manufacturer
Siemens Healthcare Diagnostics Products, LimitedModel number
IMMULITE/IMMULITE 1000 THYROGLOBULINCatalog number
LKTY1
Lot number
290
Product problems
- Adverse Event Without Identified Device or Use Problem
- Adverse Event Without Identified Device or Use Problem
Patient
Not reported
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
THE INITIAL MDR 2432235-2017-00618 WAS FILED ON 17-NOV-2017. ADDITIONAL INFORMATION (06-DEC-2017): SIEMENS HEALTHCARE DIAGNOSTICS HAS CONFIRMED THAT BIOTIN INTERFERES WITH THE TG ASSAY ON THE IMMULITE/IMMULITE 1000 PLATFORM. AT BIOTIN CONCENTRATIONS OF >5 NG/ML A NEGATIVE BIAS THAT EXCEEDS 10% HAS BEEN OBSERVED ON IMMULITE/IMMULITE 1000 PLATFORMS. THIS ISSUE AFFECTS PREVIOUSLY MANUFACTURED, IN-DATE, AND FUTURE LOTS. THE IMMULITE/IMMULITE 1000 PLATFORM INSTRUCTIONS FOR USE (IFU'S) CURRENTLY DO NOT HAVE BIOTIN LISTED AS A POTENTIAL INTERFERENT. AN URGENT FIELD SAFETY NOTICE (UFSN) IMC 18-02.A.OUS WAS SENT OUT TO OUS CUSTOMERS AND AN URGENT MEDICAL DEVICE RECALL (UMDR) IMC18-02.A.US WAS SENT TO US CUSTOMERS ON DECEMBER 13, 2017. THE UFSN AND UMDR INFORMS THE CUSTOMERS OF BIOTIN INTERFERENCE ON ALL AFFECTED IMMULITE ASSAYS. SIEMENS WILL UPDATE ALL IFU'S WITH THE APPROPRIATE BIOTIN INFORMATION.
Additional Manufacturer Narrative
THE INITIAL MDR WAS FILED ON NOVEMBER 17, 2017. FOLLOW UP MDR WAS FILED ON DECEMBER 14, 2017. CORRECTED INFORMATION (12/18/2017): BASED ON INPUT FROM THE (B)(6) RECALL COORDINATOR ON 12/18/2017 THIS RECALL WAS RE-SUBMITTED TO THE (B)(6) OFFICE (B)(4). THE CRR REPORTING NUMBER INDICATED IS INCORRECT AND WAS CHANGED TO (B)(4). AN "URGENT MEDICAL DEVICE CORRECTION (UMDC)" WAS SENT TO US CUSTOMERS ON DECEMBER 13, 2017.
Additional Manufacturer Narrative
SIEMENS IS EVALUATING THE ISSUE.
Description of Event or Problem
SIEMENS TECHNICAL OPERATIONS INTERNAL INVESTIGATION HAS IDENTIFIED THAT DISCORDANT FALSELY LOW RESULTS WERE OBTAINED ON TWO INVESTIGATIONAL SAMPLES ON AN IMMULITE/ IMMULITE 1000 INSTRUMENT WHEN USING THE THYROGLOBULIN (TG) ASSAY. THE TWO INVESTIGATIONAL SAMPLES WERE SPIKED WITH BIOTIN AT MULTIPLE CONCENTRATIONS AND RUN FOR INVESTIGATIONAL PURPOSES ON AN IMMULITE/IMMULITE 1000 INSTRUMENT. THE RESULTS OF THE TWO SAMPLES SPIKED WITH 19 NG/ML OF BIOTIN WERE FALSELY LOW WITH A BIAS OF -26% AND -32% RESPECTIVELY COMPARED TO RESULTS OF THE TWO SAMPLES WITHOUT BIOTIN. THERE WERE NO PATIENT SAMPLES IMPACTED AS THIS WAS FOR INVESTIGATIONAL PURPOSES ONLY.
Remedial action
- Recall