inforMED
MalfunctionFTT

BRAVO

Received Nov 10, 2017 · Event occurred Mar 29, 2016

Report 9710107-2017-05612 · MDR key 7019523

Device

Generic name

Camera, Still, Surgical

Model number

FGS-0312

Catalog number

FGS-0312

Lot number

29519Q

Product problems

  • Low Readings
  • Low Readings

Patient

Not reported

  • No Consequences Or Impact To Patient
  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

THE INVESTIGATION BASED ON ACCVIEW GRAPH, TOTAL TIME OF THE PROCEDURE IS 48 HOURS, THE PH VALUE AT BELOW 4 PH, AND THERE IS IGNORES IN THE PROCEDURE. NO INDICATED TO CAPSULE DETACHED EARLIER.

Description of Event or Problem

ACCORDING TO THE REPORTER, CUSTOMER CALLED TO REPORT A POSSIBLE BRAVO CAPSULE DETACHED EARLY STUDY. UPON REVIEW OF THE STUDY THE FAILURE SYMPTOM WAS DETERMINED TO BE DUE TO BRAVO LOW PH READING. A REPEAT PROCEDURE WAS NECESSARY DUE TO THE MALFUNCTION.