OMNICLOSE TROCAR SITE CLOSURE NEEDLE

Received Nov 3, 2017 · Event occurred Sep 13, 2017

Report 3007791595-2017-00012 · MDR key 6998588

Device

Generic name

Trocar Site Closure

Model number

FCL201100

Catalog number

FCL201100

Lot number

8101705035

Product problems

  • Device Operates Differently Than Expected

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

THE DHR FOR THE PRODUCT IN QUESTION WAS REVIEWED AND FOUND WITHOUT ANY IRREGULARITIES. IT IS UNCLEAR UNDER WHAT CIRCUMSTANCES COULD THE DEVICE CUT THE SUTURE CONSIDERING THE STRUCTURE AND OPERATION PRINCIPLE OF THE DEVICE. THE STRUCTURE OF THE GRASPING JAWS ARE ROUNDED WITHOUT SHARP EDGES. THE OPERATION PRINCIPLE IS TO GRASP THE SUTURE AND HOLD IT IN THE LOOP CREATED BY THE JAWS INSTEAD OF FIRMLY CLIPPING IT. A REVIEW OF THE RECORD INDICATES THAT THERE IS NO SIMILAR ISSUE SINCE THE PRODUCT IS LAUNCHED. THE EVENT WILL BE CLOSED. IF THE ADDITIONAL INFORMATION COULD BE OBTAINED FOR FURTHER INVESTIGATION, THE SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem

WHEN THE GRASPERS WERE CLOSED AND PULLING THE SUTURE IN, IT CUT THROUGH THE SUTURE INSTEAD OF GRASPING IT.