MalfunctionFTT

BRAVO

Received Nov 1, 2017 · Event occurred Oct 24, 2017

Report 9710107-2017-05598 · MDR key 6992751

Device

Generic name

Electrode, Ph, Stomach

Model number

FGS-0312

Catalog number

FGS-0312

Lot number

34582Q

Product problems

  • Failure To Adhere Or Bond
  • Failure To Adhere Or Bond

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem

ACCORDING TO THE REPORTER, THE CUSTOMER CALLED WITH A BRAVO CAPSULE THAT FAILED TO DETACH FROM THE DELIVERY DEVICE TO THE PATIENTS ESOPHAGUS. FOLLOW-UP HAS BEEN SENT FOR ADDITIONAL INFORMATION.

Additional Manufacturer Narrative

DEVICE EVALUATION: ONE BRAVO PH CAPSULE DELIVERY DEVICE AND ONE CAPSULE WERE RECEIVED FOR INVESTIGATION. THE CAPSULE WAS NOT ATTACHED TO THE DELIVERY DEVICE. VISUAL INSPECTION DID NOT REVEAL ANY DAMAGE, AND APPEARS TO HAVE FUNCTIONED WITHIN SPECIFICATION. FUNCTIONAL TESTING COULD NOT BE PERFORMED BECAUSE THIS IS A SINGLE USE DEVICE AND ONCE THE CAPSULE IS DELIVERED, IT CANNOT BE FUNCTIONALLY TESTED. INVESTIGATION CONCLUSION FOR THE FAILURE TO ATTACH COULD NOT BE RELIABLY DETERMINED. A REVIEW OF THE DEVICE HISTORY RECORD WAS PERFORMED AND INDICATES THAT THE PRODUCT WAS RELEASED MEETING FINISHED PRODUCT SPECIFICATIONS. WE WERE UNABLE TO CONFIRM THE CUSTOMER¿S REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.