ELECSYS T4 ASSAY

Received Oct 31, 2017 · Event occurred Dec 19, 2016

Report 1823260-2017-02488 · MDR key 6991011

Device

Generic name

Radioimmunoassay, Total Thyroxine

Manufacturer

Roche Diagnostics

Model number

NA

Catalog number

12017709122

Lot number

15227402

Product problems

  • High Test Results
  • High Test Results

Patient

63 YR

  • Misdiagnosis
  • Radiation Exposure, Unintended
  • Misdiagnosis
  • Radiation Exposure, Unintended

Narrative

Additional Manufacturer Narrative

(B)(4).

Description of Event or Problem

THE CUSTOMER STATED THAT THEY RECEIVED ERRONEOUS RESULTS FOR EIGHT SAMPLES FROM THE SAME PATIENT TESTED FOR THE ELECSYS TSH ASSAY (TSH), THE ELECSYS T4 ASSAY (THYROXINE), ELECSYS T3 (TRIIODOTHYRONINE), THE ELECSYS T-UPTAKE ASSAY (THYROXINE UPTAKE), THE ELECSYS FT4 II ASSAY (THYROXINE FREE), AND ELECSYS FT3 III (TRIIODOTHYRONINE FREE) ON A COBAS 6000 E 601 MODULE (E601). ALL ERRONEOUS RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE RESULTS FOR THE MENTIONED THYROID ASSAYS DID NOT MATCH THE CLINICAL PICTURE OF THE PATIENT. THIS MEDWATCH WILL APPLY TO THE THYROXINE ASSAY. (B)(6). THE PATIENT HAS A HISTORY OF MULTIPLE SCLEROSIS AND STARTED TAKING "100 MG TID" OF BIOTIN MEDICATION ON (B)(6) 2016. THE PATIENT WAS REFERRED TO AN ENDOCRINOLOGIST FOR HYPERTHYROIDISM, WITH AN INITIAL CONSULTATION ON (B)(6) 2016. IN AN ASSESSMENT PERFORMED WITH THE PATIENT ON (B)(6) 2017, THE PROVIDER NOTED THAT THE PATIENT HAD A TRIIODOTHYRONINE VALUE OF 97 NG/DL AND A THYROXINE FREE VALUE OF > 5 NG/DL. THE PROVIDER RECOMMENDED FOR THE PATIENT TO HAVE A THYROID UPTAKE EXAM IN ORDER TO DISTINGUISH GRAVES' DISEASE FROM THYROIDITIS. IN (B)(6) 2016, THE PROVIDER NOTED THAT THE PATIENT'S THYROID FUNCTION RESULTS FROM (B)(6) 2016 APPEARED TO BE QUITE HIGHER THAN PREVIOUS RESULTS. THE PATIENT HAD A THYROID UPDATE EXAM PERFORMED, WHICH REVEALED A MILDLY INCREASED RADIO-TRACER UPTAKE AT 31.8% (30% IS UPPER LIMIT OF NORMAL RANGE) AND THIS IS CONSISTENT WITH GRAVES' DISEASE. IN (B)(6) 2016 AFTER THE THYROID UPTAKE EXAM, THE PATIENT WAS PRESCRIBED 30 MG METHIMAZOLE MEDICATION EVERY DAY AS TREATMENT FOR HER HYPERTHYROIDISM. REPEAT LABORATORY TESTING PERFORMED ON (B)(6) 2017 REMAINED UNCHANGED. AFTER MEDICATION THERAPY WAS DETERMINED TO BE INEFFECTIVE, THE PATIENT AGREED TO THE NEXT STEP IN TREATMENT. THE PATIENT HAD RADIOACTIVE IODINE (RAI) ABLATION OF HER THYROID PERFORMED ON (B)(6) 2017. AFTER THE RAI ABLATION, THE PROVIDER FOLLOWED THE PATIENT'S THYROID RESULTS CLOSELY, EXPECTING HER TO BECOME HYPOTHYROIDIC. CLINICALLY, THE PATIENT WAS EXHIBITING SIGNS OF HYPOTHYROIDISM SUCH AS FATIGUE AND MEMORY LOSS. HOWEVER, THE PROVIDER DID NOT START THE PATIENT ON LEVOTHYROXINE MEDICATION TREATMENT, SINCE THYROID RESULTS ON (B)(6) 2017 CONTINUED TO INDICATE HYPERTHYROIDISM. THE PROVIDER WAS CONCERNED THAT THE THYROID VALUES WERE NOT TRENDING DOWN AS ANTICIPATED AND STARTED LOOKING INTO ALTERNATIVE EXPLANATIONS FOR THIS, INCLUDING POSSIBLE ISSUES WITH THE LABORATORY TESTS THEMSELVES. AROUND (B)(6) 2017, THE PROVIDER BECAME AWARE THAT THE PATIENT WAS TAKING THE BIOTIN MEDICATION. THE PATIENT WAS ASKED TO STOP TAKING THE BIOTIN MEDICATION AND THE PATIENT STOPPED ON (B)(6) 2017. ONE WEEK LATER ON (B)(6) 2017, THYROID TEST RESULTS MATCHED THE CLINICAL PICTURE OF THE PATIENT. THE PATIENT WAS SUBSEQUENTLY STARTED ON 125 UG OF LEVOTHYROXINE MEDICATION EVERY DAY. THE PATIENT CONTINUES TO FEEL BETTER ON LEVOTHYROXINE. THE E601 ANALYZER SERIAL NUMBER WAS (B)(4). THE CUSTOMER BELIEVES THAT THE ANALYZER IS WORKING FINE.

Additional Manufacturer Narrative

THE SAMPLES DATED (B)(6) 2017 WERE PROVIDED FOR INVESTIGATION. THE VALUES OBTAINED BY THE CUSTOMER COULD BE DUPLICATED. A VERY HIGH BIOTIN CONCENTRATION (ABOVE 700 NG/ML) WAS DETECTED IN BOTH SAMPLES. THIS CONCENTRATION IS FAR ABOVE THE ALLOWED THRESHOLD CONCENTRATION SPECIFIED IN THE METHOD SHEET FOR ALL RESPECTIVE ASSAYS. PRODUCT LABELING INDICATES THAT THERE IS NO BIOTIN INTERFERENCE UP TO 100 NG/ML FOR THE THYROXINE ASSAY. THE HIGH BIOTIN CONCENTRATION MOST LIKELY CAUSED THE HIGH THYROXINE VALUES. SAMPLES SHOULD NOT BE TAKEN FROM PATIENTS RECEIVING THERAPY WITH HIGH BIOTIN DOSES (I.E. > 5 MG/DAY) UNTIL AT LEAST 8 HOURS FOLLOWING THE LAST BIOTIN ADMINISTRATION. FOR DIAGNOSTIC PURPOSES, THE RESULTS SHOULD ALWAYS BE ASSESS IN CONJUNCTION WITH THE PATIENT´S MEDICAL HISTORY, CLINICAL EXAMINATION AND OTHER FINDINGS.