ORS-100 FLUID WARMER DRAPE

Received Oct 12, 2017 · Event occurred Sep 11, 2017

Report 8043817-2017-00028 · MDR key 6946450

Device

Generic name

Fluid Warming And Slush Drapes

Model number

ORS-100

Lot number

D171141

Product problems

  • Hole In Material
  • Hole In Material

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

A SAMPLE WAS RECEIVED FOR EVALUATION. THE SAMPLE DRAPE HAD A SMALL BURN MARK/HOLE ABOUT THREE INCHES FROM THE CENTER OF THE DRAPE. THE DHR WAS REVIEWED FOR LOT D171141 AND IT WAS NOTED THAT THIS LOT HAD (B)(4) PCS THAT WERE MANUFACTURED ON 04/24/2017. NO DEFECTS WERE REPORTED DURING QUALITY INSPECTIONS. BASED ON THE DHR AND SAMPLE REVIEW THIS ISSUE DOES NOT APPEAR TO BE THE RESULT OF A PERSONNEL, PROCESS, OR MATERIAL ISSUE. MELTS, THOUGH UNCOMMON, CAN OCCUR WHEN THE WARMER IS NOT DRAPED PROPERLY, NOT TURNED OFF BY USING THE POWER BUTTON, AND/OR THERE IS INSUFFICIENT FLUID IN THE WARMER BASIN, CONTRARY TO THE OPERATIONS MANUAL, PRODUCT LABELING, PRODUCT INSERT AND IN-SERVICE PRESENTATIONS. BECAUSE THIS APPEARS TO BE RELATED TO MISUSE BY THE CUSTOMER NO ACTIONS ARE BEING TAKEN AT THIS TIME.

Description of Event or Problem

END USER REPORTED THAT WATER WAS FOUND IN THE BASIN OF AN ORS IRRIGATION FLUID WARMER FOLLOWING THE COMPLETION OF AN ORTHO CASE. END USER STATED THAT NO SHARPS WERE INTRODUCED INTO THE DRAPE/BASIN DURING THE CASE. NO PATIENT INJURY OR TREATMENT WERE REPORTED FOR THE INCIDENT.