inforMED
InjuryJDI

11/13 S-ROM 40MM M SPEC+6

Received Oct 9, 2017 · Event occurred Sep 9, 2017

Report 1818910-2017-26181 · MDR key 6928043

Device

Generic name

Hip Femoral Head

Catalog number

136544500

Lot number

2628717

Product problems

  • Metal Shedding Debris
  • Metal Shedding Debris

Patient

66 YR

  • Foreign Body Reaction
  • Foreign Body Reaction

Narrative

Additional Manufacturer Narrative

IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem

DER STATES THAT SROM METAL ON METAL HIP IMPLANTED BY THE SURGEON APPROXIMATELY 9 YEARS AGO. ALL IMPLANTS WERE IN GOOD POSITION. PATIENT UNDERWENT REVISION SURGERY TO REPLACE THE MOM ARTICULATION. A METAL SROM HEAD OF THE SAME LENGTH WAS REIMPLANTED WITH A 40/60 + 4 10 DEGREE ALTRX LINER. THE PATIENT WAS THOUGHT TO HAVE AN ADVERSE METAL REACTION, CAUSED BY THE TRUNION OF THE STEM, THE MOM BEARING SURFACE SEEMED TO BE FINE AND NOT CAUSING ANY DEBRIS. UPDATE (B)(6) 2017 RECORDS REVIEWED. AS TRUNNION OF STEM WAS IMPLICATED AS LIKELY SOURCE OF METAL WEAR AND THE ADVERSE METAL REACTION, AN UNKNOWN S-ROM STEM HAS BEEN ADDED AND REPORTED FOR METAL BEARING WEAR AND METALLOSIS. COMPLAINT UPDATED ON: (B)(6) 2017.