inforMED
MalfunctionEAK

KENEX

Received Mar 8, 2006 · Event occurred Feb 27, 2006

Report 692325 · MDR key 692325

Device

Generic name

Suspension System, Xray Shield

Model number

350

Catalog number

*

Lot number

*

Product problems

  • Detachment Of Device Component
  • Component Falling
  • Material Disintegration

Patient

63 YR

  • Unknown (for use when the patient's condition is not known)

Narrative

Description of Event or Problem

TECHNICIAN WAS LIGHTLY TUGGING ON CEILING-MOUNTED LEADED GLASS SCREEN IN ORDER TO MOVE IT TO A DESIRABLE WORKING LOCATION. SECONDS LATER IT CAME CRASHING DOWN NEXT TO HER. THE CEILING MOUNTED STANCHION HAD FATIGUED WHERE IT SUPPORTS THE VERTICAL ARM OF THE SCREEN/ARM SYSTEM. SYSTEM WAS INSTALLED AND SUPPLIED BY SEIMENS, BUT THEY ARE NOT THE MANUFACTURER. PM WAS DONE ON 2/23/06.