inforMED
OtherNUA

THIRD WAVE CFTR INPLEX REAGENTS

Received May 10, 2006 · Event occurred Feb 15, 2006

Report 2134294-2006-00004 · MDR key 689695

Device

Generic name

Cftr Inplex Reagents

Model number

*

Catalog number

95-424

Lot number

T01AS

Product problems

  • False Positive Result

Patient

29 YR

  • Therapy/non-surgical treatment, additional
  • Therapy/non-surgical treatment, unncessary

Narrative

Description of Event or Problem

2 FALSE POSITIVE RESULTS (A HET CALL FOR THE R347P AND 3549+10KB ASSAYS) WERE REPORTED TO THE ORDERING PHYSICIANS. UPON RETESTING OF DNA FROM THE ORIGINAL SAMPLES, THE RESULTS REPORTED ON 2/15/06 WERE DISCOVERED TO HAVE BEEN FALSE-POSITIVE AND THE PHYSICIANS WERE NOTIFIED BY THE LAB USING AMENDED REPORTS AND BY TELEPHONE. THE FALSE POSITIVE RESULTS WERE USED FOR PT COUNSELING WHICH LED TO THE PHYSICIAN ORDERING A CF CARRIER SCREENING TEST ON A PT'S SPOUSE. THE LAB RECEIVED THE BLOOD SPECIMEN, BUT TESTING WAS CANCELLED BEFORE RESULTS WERE OBTAINED.

Remedial action

  • Notification