MalfunctionGET

AIR HOSE 3.0M AESC.-DRAEGER/AESC.SMALL

Received Aug 9, 2017 · Event occurred Jul 17, 2017

Report 9610612-2017-00395 · MDR key 6778758

Device

Generic name

High Speed Power Systems

Manufacturer

Aesculap Ag

Model number

GA513R

Catalog number

GA513R

Lot number

51981115

Product problems

  • Burst Container or Vessel
  • Burst Container or Vessel

Patient

Not reported

  • Hearing Loss
  • Hearing Loss

Narrative

Additional Manufacturer Narrative

EXEMPTION NUMBER: E2014018. MANUFACTURING SITE EVALUATION: EVALUATION ON-GOING

Description of Event or Problem

COUNTRY OF COMPLAINT: (B)(6). IT WAS REPORTED THAT THE AIR HOSE BURST DURING SURGERY. DUE TO THIS THERE WAS TEMPORARY HEARING LOSS BUT THE HEARING RETURNED AFTER A FEW HOURS.

Additional Manufacturer Narrative

INVESTIGATION: A DETAILED FAILURE ANALYSIS COULD NOT TAKE PLACE. THE HOSE BURST CENTRALLY. FUNCTIONALLY, THE MOTOR, THE AIR PLUG AS WELL AS THE HOSE ARE IN ORDER. A MECHANICAL DAMAGE CAN BE FOUND ON THE BURST HOSE. THE BRUISE MOST LIKELY CAUSED THE BURST OF THE AIRE HOSE. A CORRECT CIRCULATION OF THE EXHAUST AIR WAS NOT POSSIBLE. BATCH HISTORY REVIEW: THE DEVICE HISTORY FILES FOR THE MATERIAL HAS BEEN CHECKED AND FOUND TO BE ACCORDING TO THE SPECIFICATION VALID AT THE TIME OF PRODUCTION. THERE IS NO INDICATION OF A MANUFACTURING FAILURE. CONCLUSION AND ROOT CAUSE: THE MENTIONED FAILURE IS MOST PROBABLY USER RELATED. NO CAPA IS NECESSARY.