ARTHROSCOPIC SHAVERS
Received Jul 14, 2017 · Event occurred Jun 29, 2017
Report 3032391-2017-00018 · MDR key 6714611
Device
Generic name
Stryker Formula¿ Tomcat Cutter (red/white) 4.0mm, Part # 375-545-000
Manufacturer
Medline RenewalModel number
375-545-000Lot number
346255
Product problems
- Material Fragmentation
- Material Fragmentation
Patient
Not reported
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
THE DEVICE IN QUESTION WAS RETURNED FOR EVALUATION. IT WAS CONFIRMED THE DEVICE WAS REPROCESSED. THE LOT NUMBER WAS PROVIDED AND A REVIEW OF THE REPROCESSING RECORDS WAS PERFORMED. ALL PROCESSES WERE CONDUCTED AS REQUIRED PRIOR TO RELEASE. THERE WERE NO SHAVINGS ON THE RETURNED SAMPLE. FUNCTIONAL TESTS ON THE SAMPLE DETERMINED THE SAMPLE FUNCTIONED ACCORDING TO PRODUCT DESIGN AND THE SAMPLE MET PRODUCT SPECIFICATIONS. WE HAVE NOT IDENTIFIED A MANUFACTURING DEFECT TO BE A CAUSE OR CONTRIBUTING FACTOR TO THE INITIAL REPORTED ISSUE. A ROOT CAUSE COULD NOT BE DETERMINED HOWEVER, DUE TO THE REPORTED INCIDENT AND THE SUBSEQUENT NEED FOR SURGICAL INTERVENTION, THE MEDWATCH IS BEING FILED.
Description of Event or Problem
WE RECEIVED A REPORT INDICATING THAT A REPROCESSED STRYKER FORMULA¿ TOMCAT CUTTER LEFT DEBRIS IN THE KNEE. THE DEBRIS WAS REMOVED BY THE SURGEON USING IRRIGATION, AND THE PROCEDURE CONTINUED WITHOUT FURTHER INCIDENT.