inforMED
InjuryHRX

ARTHROSCOPIC SHAVERS

Received Jul 14, 2017 · Event occurred Jun 29, 2017

Report 3032391-2017-00018 · MDR key 6714611

Device

Generic name

Stryker Formula¿ Tomcat Cutter (red/white) 4.0mm, Part # 375-545-000

Manufacturer

Medline Renewal

Model number

375-545-000

Lot number

346255

Product problems

  • Material Fragmentation
  • Material Fragmentation

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

THE DEVICE IN QUESTION WAS RETURNED FOR EVALUATION. IT WAS CONFIRMED THE DEVICE WAS REPROCESSED. THE LOT NUMBER WAS PROVIDED AND A REVIEW OF THE REPROCESSING RECORDS WAS PERFORMED. ALL PROCESSES WERE CONDUCTED AS REQUIRED PRIOR TO RELEASE. THERE WERE NO SHAVINGS ON THE RETURNED SAMPLE. FUNCTIONAL TESTS ON THE SAMPLE DETERMINED THE SAMPLE FUNCTIONED ACCORDING TO PRODUCT DESIGN AND THE SAMPLE MET PRODUCT SPECIFICATIONS. WE HAVE NOT IDENTIFIED A MANUFACTURING DEFECT TO BE A CAUSE OR CONTRIBUTING FACTOR TO THE INITIAL REPORTED ISSUE. A ROOT CAUSE COULD NOT BE DETERMINED HOWEVER, DUE TO THE REPORTED INCIDENT AND THE SUBSEQUENT NEED FOR SURGICAL INTERVENTION, THE MEDWATCH IS BEING FILED.

Description of Event or Problem

WE RECEIVED A REPORT INDICATING THAT A REPROCESSED STRYKER FORMULA¿ TOMCAT CUTTER LEFT DEBRIS IN THE KNEE. THE DEBRIS WAS REMOVED BY THE SURGEON USING IRRIGATION, AND THE PROCEDURE CONTINUED WITHOUT FURTHER INCIDENT.