NEUROSENSOR COMBINED ICP/CBF PARACH PROBE, 1XL
Received Sep 9, 2005
Report 2023988-2005-00037 · MDR key 671250
Device
Generic name
Neurosensor
Manufacturer
Integra Neuro SciencesModel number
*Catalog number
NS-P
Lot number
W050708
Product problems
- No Display/Image
- Use of Device Problem
- Malposition of Device
Patient
Not reported
- Unknown (for use when the patient's condition is not known)
Narrative
Description of Event or Problem
A CBF (CEREBRAL BLOOD FLOW) MONITOR AND EVD WERE PLACED. THE SURGEON DID NOT UTILIZED A LICOX. THE CBF AND ICP WAVEFORM NEVER MATERIALIZED AT EITHER 20MM OR 25MM DEPTHS. THE NEUROSURGEON FELT IT MAY HAVE BEEN ATTRIBUTED TO THE PT'S SKULL PATHOLOGY. THE PT'S HEAD CT WAS VIEWED AND THE SKULL APPEARED VERY THICK, IT WAS ALSO OBSERVED THAT THE NEUROSURGEON PLACED THE CATHETERS MORE ANTERIOR THAN WHAT THE DIRECTIONS FOR USE (DFU) RECOMMENDS. GIVEN THE THICK BONE AND UNK DEPTH OF THE SUBARACHNOID SPACE, THE CATHETER MAY BE LOCATED IN THE PARENCHYMA. LOW CBF AND ICP VALUES WERE OBSERVED, HOWEVER, THE WAVEFORM WAS A FLAT LINE.