ELECSYS T4 ASSAY

Received Jul 10, 2017 · Event occurred Jun 20, 2017

Report 1823260-2017-01411 · MDR key 6696387

Device

Generic name

Radioimmunoassay, Total Thyroxine

Manufacturer

Roche Diagnostics

Model number

NA

Catalog number

12017709122

Lot number

188364

Product problems

  • Incorrect Or Inadequate Test Results
  • Incorrect Or Inadequate Test Results

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

THIS EVENT OCCURRED IN (B)(6). UNIQUE IDENTIFIER (UDI)#: (B)(4).

Description of Event or Problem

THE CUSTOMER COMPLAINED OF ERRONEOUS RESULTS FOR 1 PATIENT SAMPLE TESTED FOR ELECSYS T4 ASSAY (T4). IT IS NOT KNOWN IF ERRONEOUS RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE CUSTOMER COMPLAINED THAT THERE WERE "HUGE DEVIATIONS" FOR 1 SAMPLE TESTED FOR T4. NO ACTUAL COMPARISON DATA WAS PROVIDED. ONE RESULT OF 24.86 UG/DL WAS PROVIDED. IT IS NOT KNOWN IF THIS WAS AN INITIAL RESULT OR A REPEAT RESULT. THERE WAS NO ALLEGATION THAT AN ADVERSE EVENT OCCURRED. THE INSTRUMENT TYPE AND SERIAL NUMBER WAS NOT PROVIDED. THE CUSTOMER LATER PROVIDED QUALITY CONTROL (QC) RESULTS WHICH INDICATED THAT THE T4 REAGENT EXPIRED ON 01/31/2017. NO FURTHER INFORMATION WAS PROVIDED BY THE CUSTOMER. A SPECIFIC ROOT CAUSE COULD NOT BE IDENTIFIED. ADDITIONAL INFORMATION WAS REQUESTED FOR INVESTIGATION BUT WAS NOT PROVIDED. SINCE NO ADDITIONAL INFORMATION WAS PROVIDED, THE INVESTIGATION COULD NOT BE COMPLETED.