UF-1000I
Received Jul 7, 2017 · Event occurred Jun 8, 2017
Report 1000515253-2017-00030 · MDR key 6694778
Device
Generic name
Automated Urinalysis Analyzer
Manufacturer
Sysmex CorporationModel number
UF-1000ICatalog number
05366719
Product problems
- Adverse Event Without Identified Device or Use Problem
- Adverse Event Without Identified Device or Use Problem
Patient
Not reported
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
THE ARTICLE "PSEUDO-OUTBREAK OF EXTREMELY DRUG-RESISTANT PSEUDOMONAS AERUGINOSA URINARY TRACT INFECTIONS DUE TO CONTAMINATION OF AN AUTOMATED URINE ANALYZER" PUBLISHED IN THE JOURNAL OF CLINICAL MICROBIOLOGY ON JANUARY 4, 2012 WAS FOUND DURING A LITERATURE SEARCH BY A SYSMEX CORPORATION ASSOCIATE. THE ARTICLE DESCRIBED THE CONTAMINATION OF URINE SAMPLES THAT WERE ANALYZED ON THE UF-1000I URINE PARTICLE ANALYZER, AND SUBSEQUENTLY CULTURED. THE LAB STATED THAT THE CONTAMINATION DID NOT INFLUENCE THE BACTERIAL COUNTS PROVIDED BY THE ANALYZER, AND RESULTS WERE UNDER THE THRESHOLD OF PATHOLOGICAL BACTERIURIA. THIS LED TO DISCORDANT RESULTS BETWEEN THE SEDIMENT ANALYSIS ON THE UF-1000I AND THE CULTURE. THEY FURTHER STATED THAT THE PSEUDO-OUTBREAK WAS FACILITATED BY THE LABORATORY STAFF NOT OBSERVING STANDARD ROUTINE SAMPLE PROCESSING BY INCORRECTLY RUNNING THE SAMPLES ON THE UF-1000I, AND THEN USING THE SAME SAMPLE FOR CULTURES. P. AERUGINOSA SPECIES CAN FORM BIOFILMS ON PLASTIC DEVICES; HOWEVER, THE BACTERIAL COUNTS PROVIDED BY THE UF-1000I ANALYZER WERE NOT IMPACTED BY THE CONTAMINATION. THE UF-1000I FLUSHES THE HYDRAULIC LINES BETWEEN SAMPLES TO CLEAR PREVIOUSLY ASPIRATED SAMPLES FROM THE SYSTEM, AND DAILY SHUTDOWN USING BLEACH IS INTENDED TO KEEP THE ANALYZER CLEAN AND PREVENT CARRYOVER TO MEET THE SPECIFICATIONS IN THE UF-1000I INSTRUCTIONS FOR USE MANUAL. FOR RBC: 0.1 % OR LESS, OR 5.0/UL OR LESS FOR BACTERIA: 0.05% OR LESS, OR 5.0 UL OR LESS.
Additional Manufacturer Narrative
AFTER SYSMEX CORPORATION INVESTIGATION OF THE ALLEGED CROSS-CONTAMINATION DESCRIBED IN THE ARTICLE, ONLY THREE SIMILAR EVENTS WERE IDENTIFIED IN THE COMPLAINTS DATABASE. THE EVENTS REVEALED THAT THE RESULTS WERE FLAGGED, ALERTING THE OPERATOR TO POSSIBLE SAMPLE ABNORMALITY. EACH WERE RESOLVED BY ROUTINE SERVICING OF THE ANALYZER. THE OCCURRENCE RATE IS CONSIDERED VERY LOW. THE UF-1000I ANALYZER IS EQUIPPED WITH 2 PROBES, ONE DEDICATED TO MIXING THE SAMPLES AND THE OTHER FOR ASPIRATING THE SAMPLES. BOTH OF THE PROBES ARE RINSED PRIOR TO HANDLING THE NEXT SAMPLE. NO FAILED BACKGROUND COUNTS WERE REPORTED BY THE USER. NO SYSTEMIC ANALYZER ISSUE WAS IDENTIFIED AND NO CORRECTIVE ACTION IS REQUIRED.
Additional Manufacturer Narrative
THE UF-1000I IS AN AUTOMATED URINE ANALYZER FOR IN VITRO DIAGNOSTIC USE FOR URINALYSIS IN CLINICAL LABORATORIES TO SCREEN SPECIMENS CONTAINING CERTAIN UNEXPECTED OR ABNORMAL MICROSCOPIC ELEMENTS. ALL FINDINGS MUST BE CORRELATED WITH OTHER CLINICAL SIGNS AND SYMPTOMS PRIOR TO DIAGNOSIS AND TREATMENT. DECONTAMINATION WITH A BROAD-SPECTRUM DISINFECTANT AND THE REPLACEMENT OF THE TUBING WAS PERFORMED ON THE ANALYZER. THIS IS AN MDR-REPORTABLE EVENT DUE TO ADMINISTRATION OF UNNECESSARY ANTIBIOTICS TO A PATIENT BASED ON A FALSE POSITIVE RESULT FOR BACTERIA.
Description of Event or Problem
FROM (B)(6), 2010, A HOSPITAL-BASED OUTBREAK OF EXTREMELY DRUG-RESISTANT (XDR) PSEUDOMONAS AERUGINOSA WAS CONFIRMED AND AN INFECTION CONTROL ALERT WAS ISSUED. IT WAS DISCOVERED THAT A BIOFILM WAS FORMED IN THE FLUID SYSTEM OF THE ANALYZER AND MULTIPLE SPECIMENS WERE CONTAMINATED. ONE PATIENT RECEIVED THE ANTIBIOTIC COLISTIN DUE TO THE FALSE POSITIVE BACTERIA. THERE WAS NO REPORT OF PATIENT HARM DUE TO THE ADMINISTRATION OF ANTIBIOTICS.