inforMED
MalfunctionFQO

AMSCO

Received Nov 17, 2005 · Event occurred Nov 4, 2005

Report 662776 · MDR key 662776

Device

Generic name

Surgical Table

Manufacturer

Steris Corporation

Model number

3080RL

Catalog number

*

Lot number

*

Product problems

  • Break
  • Loose
  • Table, tilting

Patient

41 YR

  • Fall

Narrative

Description of Event or Problem

PATIENT HAVING SURGERY ON RIGHT LEG. TABLE TILTED TO LEFT TO ALLOW EXPOSURE FOR THE SURGEON. DURING THE SURGERY, THE TEAM HEARD VELCRO ON THE STRAP GIVE AND PATIENT SLID TO THE FLOOR. WHEN THE TABLE WAS EXAMINED AFTER THE EVENT, THE BOLT THAT SUPPORTS THE TILT MECHANISM WAS FOUND TO BE SHEARED OFF.