inforMED
MalfunctionGBW

TAUT INTRADUCERS 10/BX7.5 FR X 3.5

Received Jun 7, 2017 · Event occurred May 17, 2017

Report 3003898360-2017-00660 · MDR key 6619726

Device

Generic name

Catheter, Peritoneal

Manufacturer

Teleflex Medical

Catalog number

PI-93

Lot number

73K1600792

Product problems

  • Break
  • Break

Patient

Not reported

  • No Consequences Or Impact To Patient
  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

(B)(4). THE DEVICE HISTORY REVIEW FOR THE PRODUCT TAUT INTRADUCERS 10/BX7.5 FR X 3.5, LOT #73K1600792 INVESTIGATION DID NOT SHOW ISSUES RELATED TO THE COMPLAINT. THE DEVICE HAS NOT BEEN RETURNED FOR INVESTIGATION AT THIS TIME. TELEFLEX WILL CONTINUE TO MONITOR AND TREND RELATED EVENTS.

Description of Event or Problem

UPON USE OF INTRADUCER, DURING INSERTION OF THE CATHETER, THE VERY SMALL, CLEAR DISC ON TOP OF INTRADUCER OR SEAL FELL INTO THE PATIENT. THE CUSTOMER STATED THAT THEY MAY BE HITTING THE DISC WITH THE EDGE OF THE CATHETER AND PUSHING IT THROUGH BUT THAT IT HAS HAPPENED ON THREE SEPARATE OCCASIONS WITH THREE DIFFERENT LOT NUMBERS. NO PATIENT INJURY OR NEED TO RETURN FOR TREATMENT, THE PHYSICIAN IS SIMPLY REMOVING THE PIECE THEN AND THERE.