inforMED
MalfunctionFLF

3017 ETO STERILIZER

Received Jun 2, 2017 · Event occurred May 3, 2017

Report 3005899764-2017-00025 · MDR key 6609261

Device

Generic name

Sterilizer

Product problems

  • Device Alarm System
  • Device Operates Differently Than Expected

Patient

Not reported

  • No Consequences Or Impact To Patient
  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

A PROCEDURE DELAY WAS REPORTED DUE TO THE EVENT. A STERIS SERVICE TECHNICIAN ARRIVED ONSITE TO INSPECT THE STERILIZER AND FOUND THE VACUUM PUMP TRANSDUCER REQUIRED REPLACEMENT. THE VACUUM PUMP TRANSDUCER WAS NOT PROVIDING ACCURATE READINGS OF THE PRESSURE INSIDE THE CHAMBER SUBSEQUENTLY CAUSING THE "TOO LONG IN EVACUATION" ALARM. THE TECHNICIAN REPLACED THE VACUUM PUMP TRANSDUCER, TESTED THE STERILIZER AND CONFIRMED IT TO BE OPERATIONAL.

Description of Event or Problem

THE USER FACILITY REPORTED THEIR STERILIZER ALARMED FOR "TOO LONG IN EVACUATION". NO REPORT OF INJURY.