PROBES, DISPOSABLE SKIN TEMP, STERILE
Received May 3, 2017 · Event occurred Feb 15, 2017
Report 3010838917-2017-00031 · MDR key 6541388
Device
Generic name
Thermometer, Electronic, Clinical
Manufacturer
Deroyal Industries, IncCatalog number
M1024222
Lot number
UNKNOWN
Product problems
- Low Readings
Patient
Not reported
- Alteration In Body Temperature
- Alteration In Body Temperature
Narrative
Additional Manufacturer Narrative
(B)(4)- PER INITIAL COMPLAINT DETAILS, THE COMPLAINT SAMPLE HAS BEEN DESTROYED AND THE LOT NUMBER IS UNKNOWN. NO FURTHER EVALUATION WILL BE COMPLETED. THE REPORTED PRODUCT CODE (M1024222) HAS BEEN SET TO END OF LIFE AND IS NO LONGER MARKETED BY CAREFUSION. (B)(4).
Description of Event or Problem
DIFFERENT VALUES WHEN MEASURING WITH SKIN TEMPERATURE PROBE AND FOLEY CATHETER PROBE. THE PATIENT WAS IN ICU AND CAREFUSION SKIN PROBE WAS USED FOR TEMPERATURE MEASURING. THE PATIENTS TEMPERATURE VALUES (MEASURED FROM ARMPIT) WERE 37,1- 36,6 -35,3 BY AN HOUR. THE NURSE HAD A FEELING (WHEN TRIED PATIENT SKIN WITH THE HAND) THE PATIENT´S SKIN WAS HOT AND THEN HE/SHE MEASURED TEMPERATURE FROM THE EAR AND THE VALUE WAS 39,9. THEN PATIENT FOLEY CATHETER WAS CHANGED TO THE ONE WHICH MEASURES TEMPERATURE (COVIDIEN). THEN THE VALUE WAS 40,1°C. THE CONCLUSION WAS THAT OUR SKIN TEMPERATURE PROBE DIDN´T GIVE RIGHT RELIABLE VALUES. CUSTOMER REPORTED: ¿TEMPERATURE HAD BEEN PROBABLY DANGEROUSLY HIGH ALREADY A LONG TIME BEFORE NOTICED. THEN THEY HAD TAKEN A BLOOD CULTURES AND CONTACTED NEUROLOGIST AND INFECTION DOCTOR AND ADDED ANTIBIOTICS FOR THE PATIENT.¿