PROBNP II, N-TERMINAL PRO B-TYPE NATRIURETIC PEPTIDE
Received Apr 14, 2017 · Event occurred Mar 31, 2017
Report 1823260-2017-00788 · MDR key 6494123
Device
Generic name
Test,natriuretic Peptide
Manufacturer
Roche DiagnosticsModel number
NACatalog number
04842464190
Lot number
ASKU
Product problems
- Low Test Results
- Low Test Results
Patient
Not reported
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
SAMPLE FROM THE PATIENT WAS SUBMITTED FOR INVESTIGATION AND THE CUSTOMER¿S RESULT OF < 10 PG/ML WAS CONFIRMED. TWO POINT MUTATIONS (R72H AND E69D) IN THE EPITOPE OF THE DETECTION ANTIBODY WERE SUSPECTED. TESTING WITH THE PROBNP II REAGENT SUPPLEMENTED WITH ADDITIONAL DETECTION ANTIBODIES FOR THESE MUTATIONS SHOWED THAT THE BNP IN THE SAMPLE CONTAINED THE R72H MUTATION. THEREFORE, IT IS NOT DETECTED BY THE ELECSYS PROBNP II IMMUNOASSAY. THIS TYPE OF POINT MUTATION IS EXTREMELY RARE AND HAS ONLY BEEN FOUND IN COMPLAINT SAMPLES FROM (B)(6). A DISCLAIMER WAS ADDED TO PRODUCT LABELING.
Additional Manufacturer Narrative
UNIQUE IDENTIFIER (UDI)#: (B)(4). THIS EVENT OCCURRED IN (B)(6).
Description of Event or Problem
THE CUSTOMER QUESTIONED THE DIFFERENCE IN RESULTS BETWEEN THE ELECSYS PROBNP II IMMUNOASSAY AND ANOTHER BNP ASSAY FOR ONE PATIENT. THE PATIENT WAS MOVED FROM THE PEDIATRIC DEPARTMENT TO A CARDIOVASCULAR DEPARTMENT. THE CARDIOVASCULAR PHYSICIAN TESTED THE PATIENT'S PROBNP AND THE RESULT WAS < 10 PG/ML. THE RESULT FOR THE BNP ASSAY WAS >800 PG/ML FROM THE SAME SAMPLE. THE PHYSICIAN CHECKED THE MEDICAL HISTORY FOR THE PATIENT AND FOUND THE RESULT FOR THE BNP ASSAY MORE CLOSELY MATCHED THE CLINICAL PICTURE FOR THE PATIENT. HE ALSO FOUND THAT THE DISCREPANCY BETWEEN THE PROBNP AND BNP RESULTS HAS BEEN ONGOING FOR FIVE YEARS FOR THIS PATIENT. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER USED A COBAS 8000 E 602 MODULE. THE SERIAL NUMBER WAS REQUESTED BUT WAS NOT PROVIDED.