MEDEX¿ DOUBLE Y-CONNECTOR
Received Apr 12, 2017
Report 3012307300-2017-00901 · MDR key 6487541
Device
Product problems
- Material Disintegration
- Leak/Splash
- Loose or Intermittent Connection
- Material Disintegration
- Leak/Splash
- Loose or Intermittent Connection
Patient
Not reported
- No Consequences Or Impact To Patient
- No Consequences Or Impact To Patient
Narrative
Additional Manufacturer Narrative
TWO USED MEDEX¿ DOUBLE Y-CONNECTORS WERE RETURNED FOR INVESTIGATION. THE RETURNED DEVICES WERE RECEIVED WITHOUT PORT PROTECTORS, BUT WERE OTHERWISE ASSEMBLED ACCORDING TO THE APPROVED DRAWING. FURTHER OBSERVATION SHOWED THAT ONE VALVE WAS BROKEN INSIDE THE FEMALE LUER LOCK IN ONE OF THE DEVICES. A CAUSE FOR THE BROKEN VALVE COULD BE USING EXCESSIVE FORCE WHEN DISCONNECTING THE LINE BY USING FORCEPS, AS THE REPORTER INDICATED. THE CONNECTION BETWEEN THE VALVE AND FEMALE LUER LOCK IS SUPPOSED TO BE GLUED AS PER SPECIFICATION, WITH NO SCREWING IN AND OUT AVAILABLE. CONNECTION SHOULD BE SCREWED IN AT THE INLET SIDE OF THE VALVE. FUNCTIONAL TESTING WAS PERFORMED AND FOUND A VISIBLE LEAK FROM UNDERNEATH THE CHECK VALVE. FURTHER OBSERVATION REVEALED THIS VALVE WAS BROKEN TOO BUT IT WAS STILL POSSIBLE TO SCREW IN THE LOCK. THE COMPLAINT WAS CONFIRMED. THE ROOT CAUSE WAS DETERMINED A USER INTERFACE ISSUE. THE USER INTERFACED WITH THE DEVICE IN A MANNER THAT WAS INCONSISTENT WITH THE INSTRUCTIONS FOR USE.
Additional Manufacturer Narrative
SMITHS MEDICAL HAS RECEIVED THE SAMPLE DEVICE. A FULL EVALUATION IS ANTICIPATED, BUT NOT YET BEGUN AS THE DEVICE IS CURRENTLY IN TRANSIT TO THE INVESTIGATION SITE.
Description of Event or Problem
IT WAS REPORTED THAT A MEDEX¿ DOUBLE Y-CONNECTOR WAS ATTACHED TO A PATIENT IN THE INTENSIVE CARE UNIT FOR A NORADRENALINE INFUSION WHEN IT WAS OBSERVED THAT THE CHECK VALVE STARTED TO LEAK WITHIN 24 HOURS OF USE. IT WAS NOTED THAT FORCEPS WERE USED TO DISCONNECT THE INFUSION AND THE EXTENSION, WHICH CAUSED THE BONDING MATERIAL TO LOOSEN AND DISINTEGRATE. NO PATIENT INJURY WAS REPORTED. SEE MFR: 3012307300-2017-00834.