inforMED
MalfunctionLMB

IMAGE MANAGEMENT SYSTEM 660HD-E

Received Apr 6, 2017 · Event occurred Mar 16, 2017

Report 1643264-2017-00190 · MDR key 6464992

Device

Generic name

Device, Digital Image Storage, Radiological

Catalog number

72204668

Product problems

  • No Display/Image

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

(B)(4).

Description of Event or Problem

IT WAS REPORTED THE SCREEN WENT BLACK ON THE IMAGE MANAGEMENT SYSTEM 660HD-E. THIS OCCURRED DURING THE PROCEDURE. A BACKUP DEVICE WAS READILY AVAILABLE. THERE WERE NO DELAYS OR PATIENT INJURIES REPORTED.

Additional Manufacturer Narrative

DEVICE INVESTIGATION NARRATIVE - COMPLAINT OF BLANK SCREEN COULD NOT BE REPRODUCED. UNIT PASSED FUNCTIONAL TESTING DURING 6 HOUR BURN-IN INCLUDING TOUCHSCREEN FUNCTIONS AND CAPTURING VIDEO STILLS AND CLIPS. DISPLAY REMAINED CONSTANT THROUGHOUT FUNCTIONAL TESTING AND BURN-IN. ALL VIDEO INPUTS/OUTPUTS NORMAL AND UNIT'S SOFTWARE PERFORMED AS EXPECTED. UNIT WAS TESTED IN BOTH PAL AND NTSC FORMATS AND BOTH ANALOG AND DIGITAL VIDEO OUTPUTS. NO PROBLEM FOUND. A REVIEW OF THE DEVICE HISTORY RECORD FOR SERIAL NUMBER (B)(4) SHOWS THAT THIS UNIT PASSED ALL ACCEPTANCE CRITERIA AND WAS COMPLIANT UPON RELEASE FOR DISTRIBUTION ON FEBRUARY OF 2017. NO CONTAINMENT OR CORRECTIVE ACTIONS ARE RECOMMENDED AT THIS TIME.