inforMED
MalfunctionIOY

SUPPORT ARM 177

Received Apr 6, 2017 · Event occurred Mar 23, 2017

Report 8010042-2017-00149 · MDR key 6463301

Device

Generic name

Support, Arm

Catalog number

61481720

Lot number

2010/02

Product problems

  • Break

Patient

Not reported

  • No Patient Involvement
  • No Patient Involvement

Narrative

Additional Manufacturer Narrative

NO PARTS WERE RETURNED. INVESTIGATION WAS BASED ON A PICTURE WE HAD RECEIVED. ACCORDING TO THE PICTURE, THE SUPPORT ARM BROKE AT THE JOINT NEAREST TO THE BRACKET. THE SUPPORT ARM IS A CASTING AND IT MOST PROBABLY DEVELOPED A CRACK AT AN EARLIER OCCASION EITHER BY OVERLOADING, OR IMPACT, AND THIS LED TO THE REPORTED BREAKING. THE SUPPORT ARM HAS BEEN SUCCESSFULLY TESTED FOR MECHANICAL STRENGTH AND IS DESIGNED ACCORDING TO STANDARD. IT IS UNKNOWN UNDER WHICH CIRCUMSTANCES THE SUPPORT ARM BROKE.

Description of Event or Problem

IT WAS REPORTED THAT THE SUPPORT ARM BROKE AT THE LOWER END. THERE WAS NO PATIENT INVOLVEMENT REPORTED. (B)(4).