RESPONSE 5.5/6.0 SPINE SYSTEM
Received Mar 29, 2017 · Event occurred Mar 2, 2017
Report 3006460162-2017-00005 · MDR key 6441955
Device
Generic name
Pedicle Screw Spinal System, Adolescent Idiopathic Scoliosis
Manufacturer
Orthopediatrics Corp.Catalog number
00-1003-4001
Product problems
- Detachment Of Device Component
- Device Issue
- Detachment Of Device Component
- Device Issue
Patient
14 YR
- Congenital Defect/Deformity
- Congenital Defect/Deformity
Narrative
Additional Manufacturer Narrative
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. CORRECTIONS: PATIENT AGE, UDI, EXPLANT DATE. ADDITIONAL INFORMATION: CUSTOMER DID NOT RETURN THE EXPLANTED DEVICES TO ORTHOPEDIATRICS, CORP. CORRECTION/REMOVAL NUMBER: 3006460162-04-11-2017-001-C. A DEVICE HISTORY RECORDS REVIEW WAS UNABLE TO BE PERFORMED AS THE LOT NUMBER OF THE DEVICE INVOLVED IN THE EVENT IS UNKNOWN. INVESTIGATION RESULTS CONCLUDED THAT THE REPORTED EVENT WAS DUE TO THE SURGEON NOT ADEQUATELY FOLLOWING THE STEPS OUTLINED IN THE SURGICAL TECHNIQUE FOR "FINAL TIGHTENING" OF THE SET SCREW AND THE CONFIRMATION THAT THE ROD IS FULLY SEATED INTO THE BOTTOM OF THE PEDICLE SCREW PRIOR TO TINAL TIGHTENING. UPDATED SURGICAL TECHNIQUE, SALES REP NOTIFICATIONS AND TRAINING HAS BEEN UPDATED AND/OR DISTRIBUTED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ORTHOPEDIATRICS WILL CONTINUE TO MONITOR FOR TRENDS.
Description of Event or Problem
SCHEUERMANN'S KYPHOSIS CASE DONE (B)(6) 2016. USED ALL POLY SCREWS AND 6.0 COBALT CHROME RODS. PATIENT CAME BACK FOR 1 YEAR FOLLOW UP, AND BOTTOM 3 SCREWS ON BOTH SIDES BECAME DISENGAGED. AT TIME OF REVISION SURGERY, SALES REP NOTED THAT SURGEON STATED THAT HE HAD USED THE WRONG INSTRUMENT (JIMMY DEVICE) TO ACTUALLY REDUCE ROD INTO PEDICLE SCREWS, SO THEREFORE THEY WERE NOT TIGHTENED AS HE HAD THOUGHT.
Remedial action
- Notification