inforMED
MalfunctionFPK

MEDRAD SWABBABLE VALVE TRANSFER SET

Received Mar 22, 2017 · Event occurred Feb 24, 2017

Report 2520313-2017-00018 · MDR key 6424985

Device

Generic name

Tubing, Fluid Delivery

Manufacturer

Bayer Medical Care

Model number

60729202

Catalog number

SVTS

Lot number

UNKNOWN

Product problems

  • Device Contamination with Chemical or Other Material

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

PRODUCT ANALYSIS RECEIVED AND EXAMINED THE RETURNED PRODUCT. VISUAL EXAMINATION OF THE RETURNED PRODUCT CONFIRMED THE PRESENCE OF PARTICULATE INSIDE THE TUBING. EXAMINATION OF THE PARTICULATE FOUND IT TO BE PARTIALLY EMBEDDED WITHIN THE TUBING WALL. THE PARTICULATE MEASURES APPROXIMATELY 1.25 CM IN LENGTH. OUR INVESTIGATION DETERMINED THAT THE PARTICULATE NOTED IN THE SYRINGE WAS INTRODUCED DURING THE MANUFACTURING PROCESS. THE SYRINGE KIT INSTRUCTIONS FOR USE INSTRUCTS THE USER TO "VISUALLY INSPECT CONTENTS AND PACKAGE BEFORE EACH USE". THIS VISUAL INSPECTION IS TO ENSURE PARTICULATES ARE NOTICED AND MITIGATED, WHICH IS WHAT OCCURRED IN THIS REPORTED INSTANCE.

Description of Event or Problem

THE FOLLOWING INFORMATION WAS OBTAINED FROM THE CUSTOMER: THE CUSTOMER REPORTED BLACK FOREIGN BODY INSIDE THE TRANSFER SET TUBING. NO INJURY OR ADVERSE EVENT WAS REPORTED.