inforMED
MalfunctionCGL

I-STAT CREA CARTRIDGE

Received Mar 17, 2017 · Event occurred Mar 13, 2017

Report 2245578-2017-00011 · MDR key 6412582

Device

Generic name

Crea Cartridge

Model number

NA

Catalog number

03P84-25

Lot number

A16339

Product problems

  • False Positive Result
  • False Positive Result

Patient

67 YR

  • Breast Mass
  • Breast Mass

Narrative

Additional Manufacturer Narrative

APOC INCIDENT # (B)(4). THE INVESTIGATION WAS COMPLETED ON 04/19/2017. RETAIN PRODUCT WAS TESTED AND FUNCTIONING ACCORDING TO SPECIFICATION. RETURN PRODUCT WAS NOT AVAILABLE FOR INVESTIGATION.

Additional Manufacturer Narrative

APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Description of Event or Problem

ON (B)(6) 2017, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CREA CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT RESULT OF 2.3 ON A (B)(6) FEMALE PATIENT. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. THE CUSTOMER STATES THAT RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. DATE: (B)(6) 2017, SAMPLE: A, COLLECTED: 11:40, TESTED: 11:53, RESULT: 2.3; (B)(6) 2017, A, 11:40, 12:21, 0.7. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. THE INVESTIGATION IS UNDERWAY.