INBONE®
Received Mar 1, 2017 · Event occurred Feb 6, 2017
Report 1043534-2017-00024 · MDR key 6368320
Device
Generic name
Prosthesis, Ankle, Semi-constrained, Cemented, Metal/polymer
Manufacturer
Wright Medical Technology, Inc.Model number
200012902Lot number
1538367
Product problems
- Break
- Disassembly
- Break
- Disassembly
Patient
46 YR
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
THE PARTS WERE RETURNED. THE TOP AND MID STEM WERE RETURNED SEATED TOGETHER AND THE BASE STEM WAS RETURNED SEPARATE. VISUAL INSPECTION SHOWS NO OVERALL GROSS DEFORMATION. THERE WAS NO DAMAGE TO THE FEMALE THREADS OF THE BASE STEM OR THE MALE THREAD OF THE MID STEM. ALSO, REVIEW OF THE MID STEM THREADING DID NOT SHOW ANY DEFORMATION OR OBSTRUCTION THAT WOULD PREVENT THE BASE STEM FROM SEATING PROPERLY. THE BASE STEM WAS ABLE TO BE SEATED ON THE MID STEM WITHOUT ANY ISSUES. RADIOGRAPHIC IMAGES WERE PROVIDED AND SHOW THE BASE STEM SEPARATED FROM THE MID STEM.
Additional Manufacturer Narrative
THE DEVICE WAS NOT RETURNED FOR EVALUATION; HOWEVER FILMS WERE SUPPLIED FOR REVIEW. ANALYSIS RESULTS ARE NOT AVAILABLE AT THE TIME OF THIS REPORT. A FOLLOW-UP REPORT WILL BE SENT WHEN THE ANALYSIS IS COMPLETE. ASSOCIATED WITH REPORTS 1043534-2017-00025 AND 1043534-2017-00026.
Description of Event or Problem
ALLEGEDLY, IT WAS REPORTED THAT THE PATIENT UNDERWENT AN ANKLE SURGICAL PROCEDURE. APPROXIMATELY 8 MONTHS POST-OP, IT WAS FOUND THAT THE TIBIAL STEM IMPLANTS ARE DISASSOCIATED OR BROKEN. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.