POS COMBO 29 PANEL TYPE

Received Feb 22, 2017 · Event occurred Jan 17, 2017

Report 2919016-2017-00003 · MDR key 6349802

Device

Generic name

Microscan Dried Gram-positive Mic/combo Panels

Manufacturer

Beckman Coulter

Model number

N/A

Catalog number

B1017-209

Lot number

NOT PROVIDED

Product problems

  • Adverse Event Without Identified Device or Use Problem
  • Adverse Event Without Identified Device or Use Problem

Patient

Not reported

  • No Information
  • No Information

Narrative

Additional Manufacturer Narrative

THE FAILURE MODE FOR THE MISIDENTIFICATION COULD NOT BE CONCLUSIVELY DETERMINED WITH THE INFORMATION AVAILABLE. THE AVAILABLE INFORMATION DOES NOT REASONABLY SUGGEST THAT THE POS COMBO 29 (PC29) PANEL (LOT NUMBER NOT PROVIDED) HAS MALFUNCTIONED. CUSTOMER PANEL DATA WAS REVIEWED WITH NO INSTRUMENT OR PANEL ISSUE INDICATED. THE CUSTOMER DID NOT PERFORM REPEAT TESTING OF THE ISOLATE ON THE PC29 PANEL AND THE ISOLATE WAS NO LONGER AVAILABLE FOR ADDITIONAL TROUBLESHOOTING. DATA REVIEW INDICATED SET-UP ISSUE (I.E OVER-INOCULATION) MAY HAVE CONTRIBUTED TO THE REPORTED DISCREPANCY. BECKMAN COULTER INTERNAL IDENTIFIER FOR THIS REPORT IS (B)(6).

Description of Event or Problem

THE CUSTOMER REPORTED A MISIDENTIFICATION OF (B)(6) AS (B)(6), ON POS COMBO 29 PANEL TYPE, LOT UNKNOWN. THE CUSTOMER PERFORMED OFFLINE TEST METHODS WHICH INDICATED THE ISOLATE WAS (B)(6). STAPHAUREX TEST WAS POSITIVE, PBP2 TEST WAS NEGATIVE, VERIGENE TEST INDICATED (B)(6) POSITIVE, AND REFERENCE LABORATORY REPORTED AS (B)(6) (OX>2). DURING A FOLLOW UP CALL ON (B)(6) 2017, CUSTOMER STATED THE ISOLATE WAS ORIGINALLY REPORTED AS (B)(6), BUT AN AMENDED REPORT WAS SENT TO THE PHYSICIAN WITH THE CORRECT IDENTIFICATION (ID) OF (B)(6). PER THE PHYSICIAN, PATIENT TREATMENT WAS AFFECTED WITH THE CHANGE IN ID HOWEVER THE CUSTOMER COULD NOT PROVIDE ADDITIONAL DETAILS. CUSTOMER STATED THE PATIENT HAD BEEN DISCHARGED BUT IT WAS UNKNOWN IF PATIENT WAS DISCHARGED PRIOR TO THE AMENDED REPORT. RETEST OF THE ISOLATE IS NOT POSSIBLE AS IT WAS NO LONGER AVAILABLE. THERE WAS NO REPORT OF DEATH OR INJURY IN CONNECTION WITH THIS EVENT.