inforMED
MalfunctionCBF

SECHRIST INDUSTRIES, INC.

Received Feb 7, 2017 · Event occurred Feb 23, 2016

Report 6307045 · MDR key 6307045

Device

Generic name

Chamber, Hyperbaric

Model number

3600ER-00

Product problems

  • Device Alarm System
  • Device Operates Differently Than Expected

Patient

76 YR

  • No Consequences Or Impact To Patient
  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

NO HARM TO PATIENT. PATIENT SAFELY REMOVED FROM CHAMBER. CHAMBER#1 S/N 360196 NOT HOLDING PRESSURE AT DEPTH OF 2.4 ATA/20.58 PSIG. NO SYSTEM ALARM. ONLY NORMAL SOUNDS THAT OCCUR DURING CHAMBER OPERATION. NOTIFIED SERENA GROUP AS SOON AS DISCREPENCY DISCOVERED. MANAGER NOTIFIED. EMERGENCY CALL PLACED TO SECHRIST (B)(4). SECHRIST TO SEND SERVICE TECHNICIAN. "OUT OF ORDER" SIGN PLACED ON CHAMBER.