Received Jan 17, 2017 · Event occurred Jun 15, 2016

Report 9610806-2017-00011 · MDR key 6258928

Device

Generic name

N Latex Hcy

Catalog number

OPAX032 SMN10445973

Lot number

802907632

Product problems

  • Low Test Results
  • Low Test Results

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

SIEMENS HEALTHCARE DIAGNOSTICS HAS OBSERVED A REDUCED ONCE-OPENED STABILITY AND A REDUCED ON-BOARD STABILITY FOR THE AFFECTED LOT 802907632 THAT MAY RESULT IN ERRONEOUSLY REDUCED OR ELEVATED HOMOCYSTEINE. THIS STABILITY ISSUE MAY LEAD TO A HIGHER THAN EXPECTED LOT-TO-LOT VARIATION AND IMPAIRED PRODUCT PERFORMANCE. URGENT MEDICAL DEVICE CORRECTION (UMDC) PP17-005.A.US WAS SENT TO CUSTOMERS IN THE UNITED STATES ON JANUARY 11, 2017 AND CORRESPONDING URGENT FIELD SAFETY NOTICE (UFSN) BR-01117_OUS WAS SENT TO ALL OUTSIDE US CUSTOMERS WHO HAVE BEEN SHIPPED THE IMPACTED LOT, 802907632. THE UMDC AND UFSN TITLED "N LATEX HCY AND N LATEX HCY (BCS® XP) LOT 802907632 DOES NOT MEET ONCE-OPENED AND ONBOARD STABILITY CLAIMS" AND INSTRUCTS CUSTOMERS TO DISCONTINUE USE AND TO DISCARD THE LOT.

Description of Event or Problem

DISCORDANT LOW HOMOCYSTEINE (HCY) RESULTS WERE OBTAINED WITH LOT 802907632 ON THE BNII SYSTEM IN A COMPARISON STUDY WITH A NEWER LOT 802908369 ON A BN PROSPEC INSTRUMENT. PATIENT RESULTS WERE REPORTED TO THE PHYSICIAN. CORRECTED REPORTS WERE ISSUED WITH THE HIGHER RESULTS FROM THE BN PROSPEC INSTRUMENT. ALL CONTROLS AND CALIBRATIONS RUN WITH LOT 802907632 WERE WITHIN LABORATORY RANGES. THERE ARE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT LOW HCY RESULTS.

Remedial action

  • Recall