N LATEX HCY
Received Jan 17, 2017 · Event occurred Aug 11, 2016
Report 9610806-2017-00009 · MDR key 6258914
Device
Generic name
N Latex Hcy
Manufacturer
Siemens Healthcare Diagnostics Products GmbhCatalog number
OPAX032 SMN10445973
Lot number
802907632
Product problems
- Low Test Results
- Low Test Results
Patient
Not reported
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
SIEMENS HEALTHCARE DIAGNOSTICS HAS OBSERVED A REDUCED ONCE-OPENED STABILITY AND A REDUCED ON-BOARD STABILITY FOR THE AFFECTED LOT 802907632 THAT MAY RESULT IN ERRONEOUSLY REDUCED OR ELEVATED HOMOCYSTEINE. THIS STABILITY ISSUE MAY LEAD TO A HIGHER THAN EXPECTED LOT-TO-LOT VARIATION AND IMPAIRED PRODUCT PERFORMANCE. URGENT MEDICAL DEVICE CORRECTION (UMDC) (B)(4) WAS SENT TO CUSTOMERS IN THE UNITED STATES ON JANUARY 11, 2017 AND CORRESPONDING URGENT FIELD SAFETY NOTICE (UFSN) (B)(4) WAS SENT TO ALL OUTSIDE US CUSTOMERS WHO HAVE BEEN SHIPPED THE IMPACTED LOT, 802907632. THE UMDC AND UFSN TITLED "N LATEX HCY AND N LATEX HCY (BCS® XP) LOT 802907632 DOES NOT MEET ONCE-OPENED AND ONBOARD STABILITY CLAIMS" AND INSTRUCTS CUSTOMERS TO DISCONTINUE USE AND TO DISCARD THE LOT.
Description of Event or Problem
DISCORDANT LOW HOMOCYSTEINE (HCY) RESULTS WERE OBTAINED WITH LOT 802907632 IN A LOT TO LOT COMPARISON STUDY WITH AN ALTERNATE LOT 802905645 ON THE BN PROSPECT INSTRUMENT. PATIENT RESULTS OF THE COMPARISON STUDY WERE NOT REPORTED TO THE PHYSICIAN(S). ALL CONTROLS RUN WITH LOTS 802907632 AND 802905645 WERE WITHIN RANGE AND THAT THERE WERE NO INSTRUMENT ERRORS. THERE ARE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT LOW HCY RESULTS.
Remedial action
- Recall