MERGE ICONNECT ENTERPRISE ARCHIVE
Received Jan 11, 2017 · Event occurred Dec 12, 2016
Report 2183926-2017-00012 · MDR key 6244625
Device
Generic name
Dfc-1024 & Dfc-512 Digital Imaging System
Manufacturer
Merge HealthcareModel number
ICONNECT ENTERPRISE ARCHIVE VProduct problems
- Device Operates Differently Than Expected
Patient
Not reported
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
MERGE HEALTHCARE INVESTIGATED THE CUSTOMER'S ALLEGATION. BASED ON THE INFORMATION AND LOGGING DATA THAT WAS GATHERED, INFORMATION SUGGESTS THAT THE INCORRECT PATIENT NAME WAS SENT FROM THE IMAGING MODALITY TO THE MERGE HEALTHCARE DEVICE. MERGE HEALTHCARE (B)(4) ARCHIVE ACCEPTS DATA FROM THE MODALITIES THAT ARE SENT BY USERS. THIS ISSUE APPEARS TO BE A USER ERROR THAT OCCURED AT THE MODALITY.
Description of Event or Problem
MERGE (B)(4) CONSOLIDATES, STANDARDIZES AND ARCHIVES DICOM AND NON-DICOM IMAGES AND DATA FROM DISPARATE PACS, SPECIALTIES AND SITES. (B)(4) WORKS WITH MERGE RADSUITE WHICH IS AN APPLICATION THAT PROVIDES A MEANS TO DISTRIBUTE, DISPLAY, AND STORE DIAGNOSTIC-QUALITY MEDICAL IMAGES IN ELECTRONIC FORMAT. THE SYSTEM DISPLAYS TRADITIONAL 2D AND RECONSTRUCTED 3D RADIOLOGICAL IMAGES USING WEB-ENABLED VIEWERS OVER BOTH LOCAL AND WIDE AREA NETWORKS. ON (B)(6) 2016 A CUSTOMER REPORTED TO MERGE HEALTHCARE SUPPORT THAT THE NAME ON A PATIENT'S STUDY HAD CHANGED. THE ISSUE WAS READILY APPARENT TO THE USER DUE TO THE WORKFLOW AT THE SITE IS TO HAVE THE TECHS CHECK IN RADSUITE THAT THE PAPERWORK ATTACHED TO THE STUDY IS CORRECT. DURING THIS PROCESS THE CUSTOMER NOTICED THAT THE NAME WAS DIFFERENT THAN EXPECTED. WITH AN INCORRECT NAME APPEARING ON A PATIENT'S REPORT THERE IS A POTENTIAL FOR AN INCORRECT DIAGNOSIS AND/OR TREATMENT. HOWEVER, THE CUSTOMER DID NOT ALLEGE ANY PATIENT INJURY AS A RESULT OF THIS ISSUE. (B)(4).