VITEK® 2 GRAM NEGATIVE TEST KIT
Received Jan 6, 2017
Report 1950204-2017-00006 · MDR key 6228614
Device
Generic name
Vitek® 2 Gram Negative Test Kit
Manufacturer
Biomerieux, IncCatalog number
21341
Lot number
241391440
Product problems
- Incorrect Or Inadequate Test Results
- Incorrect Or Inadequate Test Results
Patient
Not reported
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Description of Event or Problem
A CUSTOMER IN THE UNITED STATES NOTIFIED BIOMÉRIEUX OF A MISIDENTIFICATION ASSOCIATED WITH THE VITEK® 2 GRAM NEGATIVE TEST KIT (REFERENCE 21341) INVOLVING A URINE SAMPLE. THE CUSTOMER REPORTED THE VITEK® 2 GN TEST KIT IDENTIFIED THE ORGANISM AS PROTEUS HAUSARI; HOWEVER, THE SAMPLE WAS SENT TO API WHO IDENTIFIED THE SAMPLE AS PROTEUS VULGARIS. THERE IS NO INDICATION THAT INCORRECT RESULTS WERE REPORTED AND THE CUSTOMER DID NOT REPORT NEGATIVE PATIENT IMPACT OR ADVERSE EVENTS ASSOCIATED WITH THE DISCREPANCY. CULTURE SUBMITTALS WERE REQUESTED BY BIOMÉRIEUX. AN INTERNAL BIOMÉRIEUX INVESTIGATION WILL BE INITIATED.
Additional Manufacturer Narrative
THIS REPORT WAS INITIALLY SUBMITTED FOLLOWING NOTIFICATION THAT A CUSTOMER IN THE UNITED STATES REPORTED THE OCCURRENCE OF A LOW DISCRIMINATION IDENTIFICATION OF PROTEUS PENNERI / PROTEUS HAUSERI IN ASSOCIATION WITH THE VITEK® 2 GRAM-NEGATIVE (GN) IDENTIFICATION (ID) TEST KIT (CARD). THE EXPECTED IDENTIFICATION FOR THIS API® SURVEY ORGANISM (API PROFICIENCY URINE UR-11) WAS PROTEUS VULGARIS. INVESTIGATIONAL TESTING API® SURVEY ORGANISM INCLUDED: - VITEK® 2 GN ID CARDS [TWO CARDS EACH FOR FOUR (4) CUSTOMER LOTS, ONE (1) RANDOM LOT]: TWO (2) OF THE CUSTOMER LOTS (4 CARDS TOTAL) PROVIDED A LOW DISCRIMINATION CALL OF PROTEUS PENNERI / PROTEUS HAUSERI. THE REMAINING TWO (2) CUSTOMER LOTS AND THE RANDOM LOT (6 CARDS TOTAL) PROVIDED A LOW DISCRIMINATION CALL OF PROTEUS PENNERI / PROTEUS PNEUMOTROPICA. - API® 20E: IDENTIFIED PROTEUS VULGARIS GROUP. - VITEK® MS: LOW DISCRIMINATION CALL OF PROTEUS PENNERI / PROTEUS VULGARIS. AS THE ORGANISM IS INDOLE-POSITIVE, THE VITEK® MS CALL OF PROTEUS VULGARIS WOULD BE CORRECT. THE INVESTIGATION CONCLUDED THE API® SURVEY ORGANISM HAS AN ATYPICAL BIOCHEMICAL PROFILE FOR ANALYSIS BY THE VITEK® 2 GN KNOWLEDGE BASE.